6.11 General Specifications
Item
Specifications
Standard operating
conditions
Ambient temperature:
23 ± 5°C
Ambient humidity:
55%RH ± 10% RH
Supply voltage and
frequency errors:
Within 1% of rating
Warm-up time
At least 30 minutes
Storage environment
Temperature:
–20°C to 60°C
Humidity:
20%RH to 80%RH (no condensation)
Altitude:
3000m or less
Operating environment
Temperature:
5°C to 40°C
Humidity:
20%RH to 80%RH (when the printer is not used; no condensation)
35%RH to 80%RH (when the printer is used; no condensation)
Altitude:
2000 m or less
Recommended calibration
period
1 year
Rated supply voltage
100 VAC to 240 VAC
Permitted supply voltage
range
90 VAC to 264 VAC
Rated power supply
frequency
50/60 Hz
Permitted power supply
frequency range
48 Hz to 63 Hz
Power fuse
Built in (not replaceable)
Maximum power
consumption
250 VA (when the printer is used)
Withstand voltage
(between the power
supply and case)
1.5 kVAC for 1 minute
Insulation resistance
(between the power
supply and case)
500 VDC, 10 MΩ or more
External dimensions
426 mm (W) × 266 mm (H) × 178 mm (D) when the printer cover is closed; excluding protrusions
Weight
Approx. 6.6 kg (excluding options)
Instrument cooling method Forced air cooling; outlet on the rear
Installation orientation
Horizontal or tilted using the movable leg or stand
Other orientations are prohibited.
Stacking prohibited.
Battery backup
The clock is backed up with the internal lithium battery.
Battery life: Approx. five years (at an ambient temperature of 25°C)