Working with the DL7400
Shifting the Trigger Position to the Left by Two Divisions
User’s manual section 6.2, “Setting the Trigger Position”
By shifting the waveform to the left by just 2 div, the part of the waveform after the trigger (post trigger) becomes
much easier to see.
POSITION
1
Press the POSITION key.
To enter a setting of 10%, 50%, or 90%,
press the corresponding soft key.
Trigger position mark
Pre-trigger
portion
Post-trigger portion
Trigger point
2
Turn the jog shuttle counterclockwise to set
Position to 30.0%.
The waveform shifts 2 div to the left.
Changing the Trigger Mode from Auto to Single
User’s manual section 6.1, “Selecting the Trigger Mode”
When a trigger activates in Single mode, the waveform display update stops once, and waveform acquisition stops.
Single mode is useful for observing single-shot waveforms.
2
Select Single.
1
Press the MODE key.
When waveform acquisition stops, the status changes from “Running” to “Stopped.”
MODE