IM 701450-02E
21
Working with the DL7400
Bringing the Waveform’s Entire Amplitude into View by Lowering the Vertical Position
User’s manual section 5.3, “Setting the Vertical Position of the Waveform”
2
Turn the jog shuttle counterclockwise to
set Position to –2.50 div.
1
Press the CH1 key.
Lowering the vertical position allows the entire
amplitude of the waveform to be displayed.
The ground level mark is also moved.
CH
1
Changing the Time Axis Setting from 200
µ
s/div to 100 ms/div
User’s manual section 5.12, “Setting T/div”
The time axis setting is the setting for the amount of time per grid division. If you slow (increase the value of) the time
axis setting, the display is updated according to triggers, so the instrument changes from update mode to roll mode in
which the waveform scrolls from the right to left side of the screen. Roll mode is useful when you want to observe
signals with long periods or signals that change slowly.
1
Turn the TIME/DIV knob counterclockwise to set the time axis setting to 100 ms/div.
Time axis setting on CH1
Roll mode display
TIME/DIV