4-15
IM AQ6360-02EN
Maintenance
1
2
3
4
5
App
Index
No. Message
Cause of Warning
120 USB Storage not inserted
USB storage medium not inserted.
121 USB Storage not initialized
SUB storage not initialized.
122 USB Storage is write protected
The USB storage device is write protected.
123 File not found
The specified file cannot be read because it has not been found. Or,
the file does not exist on the disk.
124 Illegal directory name
Directory cannot be created because the name is not valid.
125 Illegal file name
A file cannot be saved due to an incorrect file name.
126 Directory already exists
Directory cannot be created because a directory of the same name
already exists.
128 File is write protected
The file cannot be re-written or deleted because its property has
been set to READ ONLY.
129 USB Storage full
The file cannot be saved, because the hard disk or the USB storage
device is full.
130 Directory full
No files can be created because the directory area is full.
131 No data
There was no data to be saved, although an attempt was made to
save the file.
132 File is not a trace file
The file cannot be read because it is not a waveform file.
133 File is not a program file
The file cannot be read because it is not a program file.
134 File is not a data file
The file cannot be read because it is not a data file.
135 File is not a settings file
The file cannot be read because it is not a settings file.
138 Cannot copy
The file copy cannot be executed because the “copy from” file and
the “copy to” file have the same name.
140 No paste possible
The paste operation cannot be executed during the editing of
programs because the number of blank rows is not sufficient.
141 No merge possible
It is not possible to execute a merge during the editing of programs
because the merged result would exceed the maximum number of
rows.
142 WL calibration failed
Calibration cannot be executed because the level of a light source
is not sufficient during the wavelength calibration or because
wavelength deviations exceed the calibration range.
147 ResBW calibration failed.(Power)
Rms resolution calibration cannot be executed because the light
source level is too low.
148 ResBW calibration failed.(Spectrum)
Rms resolution calibration cannot be executed because the light
source spectrum width is too wide.
149 ResBW calibration failed.(Stability)
Rms resolution calibration cannot be executed because the light
source level is unstable.
154 This file is not compatible
The file could not be loaded because the model does not support it.
170 Illegal character
An illegal character was entered for a network name.
171 Illegal address
An illegal address was set to an IP address.
172 This version is not compatible
Attempted to load an incompatible update file.
173 Update file read error
An incorrect number of files or file corruption was found when
attempting to load the update file.
174 Invalid password
The password entered for the key lock is not correct.
175 Re-entered password is not matched
The two passwords entered for the key lock do not match.
176 Password needs to be 4-digit number
The password entered for the key lock is not a 4-digit number.
4.10 Warning Display Function
Summary of Contents for AQ6360
Page 16: ...xiv IM AQ6360 02EN ...