be aborted. And one of the following system error messages will be printed at the end:
Session aborted due to hardware failure
Session aborted by user
Session aborted (washing buffer exhausted)
Session aborted (Cleaning solution/ DI Water exhausted)
Session aborted due to thermoregulation failure
Session aborted (no empty position for priming)
Session aborted (priming failed)
Session aborted (no devices to be processed)
Warning messages
Error Message
Description
Corrective action
Cartridge Thermometer
The cartridge thermostat does not respond.
Call service.
Cartridge Temp. Adj.
Cartridge temperature is out of range.
The instrument adjusts
automatically.
Instrument Thermometer
The instrument thermostat does not
respond.
Call service.
Instrument Temp. Adj.
Instrument thermostat out of range.
The instrument adjusts
automatically.
Recoverable Errors
Error Message
Description
Corrective action
Waste Tank Full
The waste tank is full
Empty the waste tank or check
if the waste tank is connected
correctly.
Fatal Errors
Error Message
Description
Cartridge Thermometer
The cartridge thermostat does not respond.
Cartridge Temp. Adj.
Cartridge temperature is out of range.
Instrument Thermometer
The instrument thermostat does not respond.
Instrument Temp. Adj.
Instrument temperature is out of range.
The four temperature errors constitute fatal errors when the instrument is in standby.
Internal Bar Code
The internal bar code reader does not respond.
Cartridge Detection Sensor Error of detection sensor monitoring the presence of cartridges (SPS)
Mover Carriage Assembly
The dispensing frame moves wrongly.
Carousel Assembly
The tray moves wrongly.