29
V2.3.0 OM-UNN valid from 15
th
April 2015
Fig.4-2 Calibration Curve
Fig. 4-3 Curve Correction
The procedure to re-calibrating the Calibration Curve
(
Point-to-Point
)
is as follows:
The optical density of the calibrator OD
cal
will be calculated by the formula OD
cal
=
OD
450
-
OD
630;
If OD
cal0
–
TOD
cal
≤
OD
cal
≤OD
cal0
+
TOD
cal
, The calibration, OD
cal
will be accepted;
or the calibration will be cancelled and the error message will be displayed;
the K
cal
which is the correction Index will be calculated by the formula
K
cal
=
OD
cal
/ OD
cal0
;
The original Calibrator data points from the stored information from the barcode will be recalculated
according to K
cal
and the corrected curve will be stored.
The procedure to re-calibrating the Calibration Curve( 4PL ) is as follows:
OD2
OD3
OD4
OD5
OD1
C1 C2 C3
C4
C5
OD2
OD3
OD4
OD5
OD1
C1 C2 C3
C4
C5