8-4
Figure 8-1
Unscrew and remove the enzyme probes and the temperature probe from the sample module.
Remove the sample module. Remove and discard the small magnetic stir bar inside the module. Immerse the module in
the authorized disinfecting agent (see Section 8.1) for a maximum of 10 minutes. If soiling or residue is visible, the module
may be immersed in a water filled room temperature sonification bath for a maximum of 10 minutes. After cleaning, rinse
the module for 3 to 5 minutes with authorized rinsing agent (see Section 8.1). Wipe dry with a lint-free tissue.
8.4.2 Waste Module Cleaning
Unscrew the three hex screws from the top of the waste module. Disconnect the calibrator and waste tubing and remove
the waste module from the base plate.
Figure 8-2
Immerse the waste module in authorized disinfecting agent (see Section 8.1). After cleaning, rinse the module for 3 to 5
minutes with authorized rinsing agent (see Section 8.1). Wipe dry with a lint-free tissue.
Remove the O-ring from the base plate. Clean up any salt deposits or fluid on the base plate. Be sure that the base plate
and all other parts are dry.
8.4.3 Enzyme Probe Cleaning
With normal use, enzyme sensors may become fouled and cease to operate normally. A fouled sensor’s output current
will decrease and calibration may become unstable. Follow the steps below to clean the probes.
Waste Module
Allen Screws
Thumbnuts
Sample module
Enzyme Probe
Enzyme Probe
Summary of Contents for YSI 2500
Page 1: ...USER MANUAL 525021 YSI 2500 Biochemistry Analyzer OPERATIONS AND MAINTENANCE MANUAL...
Page 34: ...5 11 8 The Stat sample results are displayed on the Run Stat tab Stat sample results...
Page 88: ...13 1 13 Notices Declaration of Conformity 13 1...
Page 96: ...17 1 17 Appendix D Line Power Cord and Plug Wiring United Kingdom United States Europe...