Figure 8-6
Pumps – Right side
Figure 8-7
8.4.6 Pump Tubing Replacement
Tubing life depends on instrument usage. The buffer and calibrator pump tubing should be replaced at least every 6
months or 1000 hours sample ready.
NOTE:
The buffer pump tubing and calibrator pump tubing each require a different type of grease. It is important to apply
the correct grease to each type of tubing.
Hex Screws
Buffer Pump
BLUE
Cal Pump
RED
Sipper Pump
Pump Cover
(right side)