4-4
7. Touch [Yes] to confirm and disable the safety interlocks.
8. Touch [X] in the top left of the screen to return to the main display.
Align Sipper
4.4
It is very important that the sipper be accurately adjusted.
WARNING: Keep your hands clear of the sipper while the instrument is in operation.
1. Touch the
Service
icon.
2. From the Sipper tab of the Service screen, touch [Home].
3. Once the sipper has moved to the home position, touch [Module 1].
4. The Select Location screen will appear. Select [Module 1]. The sipper will move to sample module 1 and should
be centered above the cone shaped opening in the top of the module.
If the sipper does not move, make sure the packing material was removed.