Commissioning
TriOxmatic
®
700 (SW)
52
ba25307e06
07/2015
3.3
Commissioning / Getting the instrument ready for
measuring
Note
For the sensor to polarize, the following conditions must be fulfilled:
The sensor is connected to the measuring transmitter
The measuring transmitter is operating
Precision
measurements
Recommendation: In order to carry out precision measurements, leave
the sensor to polarize for a longer period of time, e.g. overnight, and
recalibrate it the following day.
1
Pull the protective cap off the protective hood of the sensor.
2
Leave the sensor to lie in the air for at least 60 minutes while it
is switched on (polarization).
3
Calibrate the sensor (see operating manual of the measuring
transmitter).