Maintenance, cleaning, disposal, and replacement
TriOxmatic
®
700 (SW)
74
ba25307e06
07/2015
Variant 2:
Cleaning the silver
counter electrode in a
beaker
Caution
The reference electrode must not come into contact with the
cleaning solution under any circumstances. This could destroy
the reference electrode and cause the sensor to become
defective.
1
Wipe the silver counter electrode with a lint free paper towel
and carefully remove any loose deposits.
2
Fill a beaker (150 ml, high-sided form) with approx. 25 ml RL-
AG/Oxi cleaning solution.
3
Clamp the sensor in a stand.
Submerse the electrode unit in the RL-AG/Oxi cleaning
solution to just above the silver counter electrode.