Maintenance, cleaning, disposal, and replacement
TriOxmatic
®
700 (SW)
66
ba25307e06
07/2015
Precision
measurements
Recommendation: In order to carry out precision measurements, leave
the sensor to polarize for a longer period of time, e.g. overnight, and
recalibrate it the following day.
Note
In the following cases, you must refill the sensor once more:
If there are large air bubbles
If there are air bubbles on the gold working electrode
For measurements at high water pressure, even if small air bubbles
are present. Otherwise, the membrane and, as a result, the
measuring characteristics could change.
If the sensor cannot be calibrated.