What to do if...
TriOxmatic
®
700 (SW)
80
ba25307e06
07/2015
Measured value
fluctuating heavily
Incorrect temperature
display
Mechanical damage to
the sensor
Cause
Remedy
– Membrane head loose
– Screw the membrane cap
tight
– Membrane does not fit snugly on
the gold working electrode
– Change the membrane cap,
then recalibrate (see section
5.3 and section 4.2)
Cause
Remedy
– Temperature sensor defective
– Return the sensor
Cause
Remedy
– Return the sensor