TriOxmatic
®
700 (SW)
Technical data
83
ba25307e06
07/2015
7.2
Measurement conditions
Measuring range
Polarization time
Temperature range
Signals for sensor
monitoring
(SensCheck function)
Pressure resistance
Sensor including connection cable:
The sensor meets all requirements according to article 3(3) of 97/23/
EG ("pressure equipment directive").
Type of protection
Depth of immersion
min. 10 cm; max. 100 m depth
Operating position
Any, if the minimum approach flow is guaranteed
Approach flow
≥
5 cm/s (at 1 % measurement accuracy)
D. O. concentration
0.0 ... 60.0 mg/l
D. O. saturation
0 ... 600 %
(
≈
0 ... 1200 mbar pO
2
)
In the case of re-commissioning or change of
electrolyte:
At least
60 minutes
In the case of short polarization interruptions
(depending on the duration of the interruption):
15 to 60 minutes
Measuring medium
0 °C ... + 50 °C
Storage/transport
- 5 °C ... + 50 °C
TriOxmatic 700
TriOxmatic 700 SW
SensReg
(electrolyte solution depleted)
Yes
Yes
SensLeck
(membrane cap leaking)
Yes
No
Max. allowed overpressure
10
6
Pa (10 bar)
Sensor including connection
cable
IP 68, 10 bar
Connector plug
IP 65