62
EN 5.10 SuSix Manual 2007
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
mA Output
When a
SuSix®
is connected to a power supply for the first time, the
mA output is automatically set to provide 4 mA at zero and 20 mA at the
theoretical maximum value.
Changes in the mA setting will not affect the relay output settings.
Note: Both values can be set in the range 10 % to 120 %, making it
possible to increase or decrease the mA signal.
Averaging
Units
mA output
Converter Setup
Quality improvement
No
Yes
Back Select
>
20 mA output
20 mA =
20.00 g/l
Back Select
OK
Units
NTU (Turbidity)
FTU (Turbidity)
FNU (Turbidity)
EBC (Turbidity)
% (Suspended solids)
mg/l (Suspended solids)
g/l (Suspended solids)
Kg/l (Suspended solids)
Back Select
OK
Fortsætter på side 5
Sensor Calibration
Averaging
Units
mA output
Cleaning
Solid curve
High Alarm
Low Alarm
Sensor error
Status
Factory settings
Service menu
Back Select
OK
4mA output
4 mA =
00.00 g/l
Back Select
OK
accordingly
Sensor Calibration
Yes
Use mA output
Yes
No
Back Select
OK
Averaging
0001 sec
(from 1 sek to 120 min)
Back Select
>
Yes
Adjust calibration
Select point to adjust
1 point Done
2 point Adjust
3 point
Back Select
OK
Calibration point 1
Type in lab. Value
10,2 g/l
Ok (/ Next)
Adjust
Back Select
Ok
Adjust
Calibration point 1
010,230 g/l
Back Select
>
Adjust
1
point
Adjust
Calibration point 1
0,00 g/l
Back Select
>
Adjust samples
Adjust
1. point
00010,2 g/l
Back Select
>
Adjust sample
1 Measured
10,2 g/l
2 Measured
8,11 g/l
3 Measured
5,87 g/l
Back Select
OK
Next
Select point
Adjust
2. point
00008,11 g/l
Back Select
>
Adjust
3. point
00005,87 g/l
Back Select
>
Sensor Calibration
Measurement samples
Adjust samples
Clear ”Measurement sample”
Clear ”Adjust samples”
Back Select
OK
1./2./3. point
Start sample?
Progress
[ 0% ]
Exit Start
Measurement samples
Select sample
1. point Measured
2. point Measured
3. point Measured
Back Select
OK
Clear ”Adjust samples”
Clear ”Adjustment sample”
No
Yes
Back Select
Ok
Start point
2. point
1. point
3. point
Next
Ok
Ok
Ok
Select point
Select point
1./2./3. point
Measureing
Progress
[||||||||||||||||| ] 45%
Cancel
Measure point 1/2/3
Done
Progress
[ 100% ]
Exit Next
Select point
Clear ”Measurement sample”
No
Yes
Back Select
Ok
Clear ”Measurement sample”
3 - Converter setup 1
Page3
The mA output is an active output, and the maximum load is 800 Ohm.
The upper limit for the mA output is 20,5 mA
3,75 mA indicates the mA output is not in use
3,5 mA indicates values under 4 mA level
20,5 mA indicates value higher than 20 mA level.