61
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
EN 5.10 SuSix Manual 2007
2. Select the digits one by one with the left/right keys and set the value
with the up/down keys.
3. Press OK to save the final setting.
Units
The "Units" option sets the unit for turbidity and suspended solids. The
available units are shown below.
Averaging
Units
mA output
Converter Setup
Quality improvement
No
Yes
Back Select
>
20 mA output
20 mA =
20.00 g/l
Back Select
OK
Units
NTU (Turbidity)
FTU (Turbidity)
FNU (Turbidity)
EBC (Turbidity)
% (Suspended solids)
mg/l (Suspended solids)
g/l (Suspended solids)
Kg/l (Suspended solids)
Back Select
OK
Fortsætter på side 5
Sensor Calibration
Averaging
Units
mA output
Cleaning
Solid curve
High Alarm
Low Alarm
Sensor error
Status
Factory settings
Service menu
Back Select
OK
4mA output
4 mA =
00.00 g/l
Back Select
OK
accordingly
Sensor Calibration
Yes
Use mA output
Yes
No
Back Select
OK
Averaging
0001 sec
(from 1 sek to 120 min)
Back Select
>
Yes
Adjust calibration
Select point to adjust
1 point Done
2 point Adjust
3 point
Back Select
OK
Calibration point 1
Type in lab. Value
10,2 g/l
Ok (/ Next)
Adjust
Back Select
Ok
Adjust
Calibration point 1
010,230 g/l
Back Select
>
Adjust
1
point
Adjust
Calibration point 1
0,00 g/l
Back Select
>
Adjust samples
Adjust
1. point
00010,2 g/l
Back Select
>
Adjust sample
1 Measured
10,2 g/l
2 Measured
8,11 g/l
3 Measured
5,87 g/l
Back Select
OK
Next
Select point
Adjust
2. point
00008,11 g/l
Back Select
>
Adjust
3. point
00005,87 g/l
Back Select
>
Sensor Calibration
Measurement samples
Adjust samples
Clear ”Measurement sample”
Clear ”Adjust samples”
Back Select
OK
1./2./3. point
Start sample?
Progress
[ 0% ]
Exit Start
Measurement samples
Select sample
1. point Measured
2. point Measured
3. point Measured
Back Select
OK
Clear ”Adjust samples”
Clear ”Adjustment sample”
No
Yes
Back Select
Ok
Start point
2. point
1. point
3. point
Next
Ok
Ok
Ok
Select point
Select point
1./2./3. point
Measureing
Progress
[||||||||||||||||| ] 45%
Cancel
Measure point 1/2/3
Done
Progress
[ 100% ]
Exit Next
Select point
Clear ”Measurement sample”
No
Yes
Back Select
Ok
Clear ”Measurement sample”
3 - Converter setup 1
Page3
1. Press the up/down keys to highlight the required unit and then press
OK to select it.