54
EN 5.10 SuSix Manual 2007
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Password
A password provides (and prevents) access to all the settings in the
Display and Converter unit. The code consists of a numeric 5-digit code
between 0 and 65535. If your current password is lost or forgotten, the
password protection can be overruled with the code "01750".
1. Press the up/down keys to highlight the required menu line
(here: Password) and then press OK to select it.
Specify Main Screen
Data Logger
Password
Set Sensor Name
Converter Setup
Display Setup
Back Select
OK
Login
Activate/deactivate
Change password
Back Select
OK
Set log interval
0010 sec
Back Select >
Data Logger
Set Sensor Name
SuSix1
Back Select
>
Set Sensor Name
Select line
Linie 1 - Sensor name
Linie 2 - Meas.
Linie 3 -
Linie 4 -
Linie 5
Back Select
OK
Specify Main Screen
Converter Setup
Language
Set Clock
Modbus
Factory settings
Display SW version
Back Select
OK
Display Setup
Sensor Calibration
Averaging
Units
mA output
Cleaning
Solid curve
High Alarm
Low Alarm
Sensor error
Status
Factory settings
Service Menu
Back Select
OK
Select line content
Not in use
Meas.
Homogeneity
Solid curve
Clock
Back Select
OK
Password
Type password
0000
Back Select
>
Change password
0000
Back Select
>
Activate
Deactivate
Back Select
OK
Login
Activate/
Deactivate
Change password
SuSix1 [X]
Meas.
FTU
Select Setup
Main Menu
Converter Top Display
SuSix1 [X]
Meas.
FTU
Select Setup
Converter Top Display
Specify Main Screen
Data Logger
Password
Set Sensor Name
Converter Setup
Display Setup
Back Select
OK
Main Menu
SuSix1 [X]
Prim.
mg/l
Second 0.1%
Select Setup
1 - Main menus
Page1
2. Press the up/down keys to highlight the required option and then press
OK to select it.
Specify Main Screen
Data Logger
Password
Set Sensor Name
Converter Setup
Display Setup
Back Select
OK
Login
Activate/deactivate
Change password
Back Select
OK
Set log interval
0010 sec
Back Select >
Data Logger
Set Sensor Name
SuSix1
Back Select
>
Set Sensor Name
Select line
Linie 1 - Sensor name
Linie 2 - Meas.
Linie 3 -
Linie 4 -
Linie 5
Back Select
OK
Specify Main Screen
Converter Setup
Language
Set Clock
Modbus
Factory settings
Display SW version
Back Select
OK
Display Setup
Sensor Calibration
Averaging
Units
mA output
Cleaning
Solid curve
High Alarm
Low Alarm
Sensor error
Status
Factory settings
Service Menu
Back Select
OK
Select line content
Not in use
Meas.
Homogeneity
Solid curve
Clock
Back Select
OK
Password
Type password
0000
Back Select
>
Change password
0000
Back Select
>
Activate
Deactivate
Back Select
OK
Login
Activate/
Deactivate
Change password
SuSix1 [X]
Meas.
FTU
Select Setup
Main Menu
Converter Top Display
SuSix1 [X]
Meas.
FTU
Select Setup
Converter Top Display
Specify Main Screen
Data Logger
Password
Set Sensor Name
Converter Setup
Display Setup
Back Select
OK
Main Menu
SuSix1 [X]
Prim.
mg/l
Second 0.1%
Select Setup
1 - Main menus
Page1