Turbidity and Suspended Solids Transmitter
EN 5.10 SuSix Manual 2007
Sensor
Error
- General sensor error that
may also be transferred to a relay
output.
Contact MJK.
Sensor communication
Error
- Check the cable condition
and tighten the connectors.
Sensor hardware
Error
- Change the sensor hard-
ware.
Sensor setting
Error
- Some settings are wrong or
have been changed.
Sensor supply
Error
- The sensor supply is too
low or too high. Check the sensor
cable connection.
Sensor wiper
Error
- Make sure that the wiper is
free to move and not obstructed by
dirt deposits.
System
Error
- Internal VCC error. See the
event log for details.
System error
Error
- The system software is in-
operable including faulty alarm and
sensor bits.
Contact MJK.
Turbidity offset
Error
- A zero-point probe prob-
lem. The sensor needs a zero
calibration.
Contact MJK.
Warnings
Needs higher meas.sample
Warning
- Calibrate at a higher
concentration.
Needs lower meas.sample
Warning
- Calibrate at a lower
concentration.
Sensor range exceeded
Warning
- The measured value is
out of range. Check the measuring
range and the measuring units.