Clean the exterior of the device with a non-aggressive cleaning fluid or detergent
to remove all inequities.
•
Allow to air dry and inspect for damage or deterioration of the surfaces.
•
Do not use any abrasives as this will damage the surface of the device.
•
When the device is used on a regular basis please keep it connected to mains to
reload the batteries
•
Drain, disinfect and decalcify the water reservoir on a regular basis, see sections
4.2 and 4.3, to guarantee optimal performance of the device.
1.
Wear protective gloves and goggles during the disinfection process.
2.
Create 2 liters of 0.5% Chloramine-T solution; follow instructions of the disinfectant
manufacturer (for example: 4 DISIFIN tablets in 2 liters of demineralized water
(www.disifin.co.uk)).
3.
Drain the water from the thermo unit and water tubing using the 3-way tap (29).
After drainage close the 3-way tap.
4.
Clean the white connectors (26), water tubing coupler (27), 3-way tap and lid of
thermo unit reservoir (15) using a surface disinfectant.
5.
Close the water circuit.
6.
Add 2 liters 0.5% Chloramine-T solution to the thermo unit reservoir.
7.
Connect sensors:
•
Connect the flow sensor, temperature sensors and pressure sensor cable to
the PV pump unit.
•
Submerge the flow sensor in a cup with tap water.
•
Connect separate pressure sensor e.g. from a used disposable set or new
standard pressure sensor (e.g. Edwards TruWave).
Note: No need to connect a complete disposable system!
8.
Connect power cable of the device to mains.
9.
Power on the PV pump unit.
10. Skip through setup procedure by pushing the OK-button on the PV pump unit.
Push until the display shows ”pressure zeroing”.
11.
Circulate disinfection solution
for 30 minutes at room temperature; check if the
red flow indicator wheel (28) is spinning to ensure flow.
12. Power off PV pump unit and drain thermo unit and water tubing (
see step 3
).
13.
First Rinse
: Add 2 liter of demineralized water to the thermo unit, circulate the
water for 5 minutes at room temperature; check if the red flow indicator wheel is
spinning to ensure flow.
(
follow steps 9 and 10 to start the circulation)