Figure 12: Connecting the sensors
Figure 13: Connection of flow sensor to disposable system
Every sensor is different and is not interchangeable. Connect the sensors to
the right sub connectors, else it could cause damage.
Do not spill any fluid on the electronic connectors; this may cause a deviation
of the measured values or to an alarm.
Do not load more than 15 Kg on the table top, including organ and liquids.
Open the flow sensor by pushing the aluminum clip to release the lid.
Open the lid and clamp the sensor around the silicon tube.
Push the lid to close the sensor.
Make sure that the arrow on the flow sensor is facing the same direction of the flow
through the tube (towards the reservoir). Wrong connection of this sensor will not give a
correct flow measurement.