Before placing the pump head to the magnetic pump coupling; remove the metal
clip, marked with an orange ‘remove before use’ label.
•
Connect the pump head to the magnetic pump coupling located on the outer sides
of the Liver Assist pump units (figure 3, item 14). Push the pump head in the coupling
and turn the pump head to lock it, see figure 11, ensure it is placed correctly. The
outlet of the pump heads shall be orientated horizontal to easily de-air the pump
heads.
Figure 11: Connection of pump-head
•
Connect the oxygenators to the oxygen supply using the green tube with the
incorporated gas filter.
•
Connect the thermo water tubing to the oxygenators using the water connectors
(26).
Gas supply:
It is preferable to make use of the gas supply of the operating theatre. Consult the
instructions of the oxygenator for requirements of the oxygen supply.
When there is no oxygen/gas supply available a cylinder can be used. When a cylinder is
used always check if there is enough gas available in the cylinder. XVIVO is not responsible
for incorrect use of the gas supply.
•
Place the pressure sensors to each side of the table (36).
•
Connect the pressure sensor cables to the pump units (12) and to the pressure
sensors.
•
Connect temperature sensors (10,11) according to the color coding (figure 12)
•
Connect the HA T1 sensor (red) to the HA oxygenator outlet and the PV T1 sensor
(red) to PV oxygenator outlet (figure 9 and 14).
•
Connect the PV T2 sensor (blue) to the PV perfusion line.
•
Connect the flow sensor (13).
•
Make sure to connect the sensor connection with the red dot facing upwards.
Push
Turn
The Liver Assist should not to be used in contact with flammable agents, gases
or liquids and not to be used in an oxygen rich environment.