11.01.601.7c EN
Page 20 of 40
•
Remove cap on top of the pressure sensor (transducer) and place a sterile syringe
on the opened port (figure 15).
•
Pull the blue snap tab while aspirating with the syringe until the pressure line is de-
aired and there is enough perfusion solution in the syringe.
•
Remove the syringe and replace the cap. Turn the valve on the pressure sensor
(transducer) counterclockwise to open the sensor to measure atmospheric
pressure and remove the cap (see figure 15)
•
Press the OK-button to zero the sensor to atmospheric pressure. This step will take
a few seconds.
•
Turn the pressure sensor valve back to measuring (end) position (clockwise, see
figure 15).
•
Re-place the cap
Figure 15: Priming and zeroing of the pressure transducer
Pressure zeroing
Turn transducer
valve
Press to continue
Pressure zeroing
Turn transducer
valve
Press to continue
HEPATIC ARTERY
PORTAL VEIN