11.01.601.7c EN
Page 13 of 40
Perfusion pump:
Rotary pump, pulsatile 60 BPM (HA) + continuous (PV)
Perfusion flow:
HA: up to 570 ml/min @ 12 °C / up to 1 L/min @ 37 °C
PV: up to 500 ml/min @ 12 °C / up to 2 L/min @ 37 °C
Perfusion pressure:
HA: up to 50 mmHg @12°C / up to 90 mmHg @ 37 °C
PV: up to 11 mmHg @12 °C / up to 16 mmHg @ 37 °C
Perfusion temperature:
12 °C - 37 °C
Accuracy:
Pressure: ± 1.5% or 1 mmHg
Flow: ± 0.07 L/min
Temperature: ± 1°C
Perfusion solution:
Any certified machine perfusion preservation solution (2-4 L)
Displayed:
Perfusion time, flow, pressure, temperature, reservoir
temperature, vascular resistance, menu, messages
Alarm:
Alarm sound level pressure: 65 dB(A)
Battery capacity:
20 minutes
(Lithium-ion Battery, 11.25V / 8850mAh / 99.6Wh)
Battery charging:
Self-charging if connected to Mains (min. 8 h)
Power:
AC 110V/60 Hz or 230V/50 Hz 880 VA
Fuses (HA/PV):
Littlefuse: 0215002.txp 2AT 250V HBC
Fuses thermo unit:
Littlefuse: 0215008.txp 8AT 250V HBC
Maximum load on table top:
15 kg, including organ and perfusion solution
Transport & storage
conditions:
5 - 40°C, 30 - 85 %RH
Atmospheric pressure: 50,0 kPa to 106,0 kPa
Operating conditions:
Room temperature 18-24°C, 30 - 85 %RH
Atmospheric pressure: 70,0 kPa to 106,0 kPa
Background noise level: < 50dBA
Do not use the device in a poorly ventilated area
Product life time:
7 years
Dimensions:
1120 mm x 925 mm x 625 mm
Weight:
81 kg
Ingress protection:
IP20