Table 3. Guidance and manufacturer’s declaration – RF wireless communication equipment Immunity
The Liver Assist is intended for use in the electromagnetic environment specified below. The customer or the user of this
device should assure that it is used in such an environment.
Test
frequency
(MHz)
Band
(MHz)
Service
Modulation
Maximum
power
(W)
Distance
(m)
Immunity
test level
(V/m)
Compliance
level
(V/m)
385
380 - 390
TETRA 400
Pulse
modulation
18 Hz
1,8
0,3
27
27
450
430
– 470
GMRS 460,
FRS 460
FM
± 5 kHz
deviation
1 kHz sine
2
0,3
28
28
710
704 - 787
LTE Band
13, 17
Pulse
modulation
217 Hz
0,2
0,3
9
9
745
780
810
800
– 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18 Hz
2
0,3
28
28
870
930
1720
1700
– 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
217 Hz
2
0,3
28
28
1845
1970
2450
2450 - 2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
217 Hz
2
0,3
28
28
5240
5100 - 5800
WLAN 802.11
a/n
Pulse
modulation
217 Hz
0,2
0,3
9
9
5500
5785
NOTE : The frequencies and services listed are representative examples that are based on RF wireless communications
equipment in use at the time of publication of IEC 61000-4-3. The test specification does not attempt to cover every
frequency and service used in every country.