Chapter 3 Controller Descriptions and Operation 

get the desired value. Press 

 to confirm and exit the 

SEt

 screen or exit 

automatically in 4 seconds. 

3.3 Constant temperature control 

1)

 

Connect the power by inserting the power plug, or engage the circuit breaker on the switch 
board. 

Turn on power switch and the controller runs self-check to display: 

88888

, LED on, software 

version number, 

AoFF

 (

Aon

 will show up if 

Auto

 is set to 

Aon

) settings for 

Ht

Lt

. Refer to 3.5 for 

procedures about 

Aon

 and 

AoFF

 setting. The software version number may change with 

technique updates.  

2)

 

Check and/or set the working temperature, refer to item No.2 in the table. 

3)

 

If the equipment is in the standby state, press 

 for more than 2 seconds, the circulating 

pump will start to operate and the heater is on if the setting is higher than the actual 
temperature. 

4)

 

When the temperature in the bath is lower than the set value, heating LED is on and the 
equipment is in operation and start to heat. When the set temperature and actual 
temperature are quite different, heating LED will be steadily lit with full power heating and the 
temperature will go up with full speed. As the temperature approaches the set point, the LED 
flashes as desired and the controller decreases the power for the heater. With the actual 
temperature equals to the set temperature, the controller keeps the temperature by 
constant power output. 

5)

 

To terminate the temperature control, press 

 for more than 2 seconds. The circulating 

pump and heater are stopped. The LEDs go off, and the display shows 

oFF

 

Connect the power from behind the equipment. Plug in and remove the cable by directly 
holding the plug. No dragging of cable in any part. Protect the cable from being damaged by 
contacting with hot surface of the equipment or mouse etc! 

 

Caution, Hazard of overflowing or low level! 

Never attempt to initiate the equipment if no heat transfer liquid is filled or filling volume is lower 
than the minimum level as specified; Never move the equipment if the bath tank is fully filled 
with the heat transfer liquid; The liquid level may change with the heating and cooling of the 
equipment, and with the immersion or removal of the specimen! 

Summary of Contents for UXP 8011-HC090

Page 1: ...SHANGHAI 200032 CHINA TEL 86 21 6443 1641 FAX 86 21 6403 9005 EMAIL BC XTEMP USA CN XTEMP ENVIRONMENTS SHANGHAI CO LTD ADD ROOMS 909 11 NO 2777 ZHOUHAI ROAD PUDONG NEW DISTRICT SHANGHAI 200137 CHINA T...

Page 2: ...it 14 2 2 5 Accessory cooling coil Optional 15 2 2 6 Put into operation 15 Chapter 3 Controller Descriptions and Operation 17 3 1 Controller panel 17 3 1 1 Display for temperature and working paramete...

Page 3: ...08 2 V211119 UXP8011 CD6400 CLASS I NFL RT 5 100 C Contents C 4 2 Regular check of the STB 27 4 3 Checking and replacing the fuse 28 4 4 Clear error warning and alarm 29 Chapter 5 Equipment Overview a...

Page 4: ...trical energy The operator is largely protected through the application of the appropriate standard specifications Further hazard sources may arise from the type of material being treated e g mixing d...

Page 5: ...he conditions that the equipment is operated and serviced exclusively in accordance with its intended purpose and as described in these Operating Instructions the equipment is not modified only origin...

Page 6: ...egrade electric safety of the equipment Note the effect of the altitude on the boiling point of the heat transfer liquids 1 2 Safety warning When heating and conditioning with specimen or application...

Page 7: ...clear of any obstacles Keep the equipment away from hot air emitting source direct sunshine strong magnetic field and electric sparks Neither contamination of high concentration dust or corrosive gas...

Page 8: ...iquids Use special care when using with these flammable liquids including the proper setting of the adjustable overheating protection device The circulation of the heat transfer liquid is important fo...

Page 9: ...kink and other mechanical stress develops during normal use which otherwise may result in thermal and or liquid hazards 2 Flexible hose fitting and coupling are important for the safe operation of the...

Page 10: ...azard 2 Don t touch the frozen surface or any parts with cold residual liquid Warning Hazard of frostbite Caution Hazard of possible electric shock when removed and placed in horizontal or up side dow...

Page 11: ...servicing or repairing Accessories lists UXP8011 HC090 Part Number Qty Descriptions Remarks DSGO0708 2 1 Operating instructions This document ERSG0202 1 Fuse 250V 15A D5L20 Breaking capacity 1500A BJG...

Page 12: ...f the equipment with protective ground PE For possible wet location conditions where cycle of evaporating and condensing overflowing or spillage may occur and for protection against hazard of possible...

Page 13: ...ment Use only hose fitting and coupling provided with the equipment or recommended by Xtemp The ratings of pressure and temperature for the hose fitting and the jacket container or heat exchanger of t...

Page 14: ...stacles so that no close loop of the air circulating may develop If no external circulating is necessary use hose fitting and coupling to short circuit the liquid connections as for an external applic...

Page 15: ...ne adaptor SL The connection to the pump outlet and external circulating kit shall be hydraulically tight Loose connection may result in hazard of liquid spray The pump inlet shall be clear of any obs...

Page 16: ...coil Optional Fig 5 Cooling coil For immersion heating bath GP series and circulators completed with open bath tank the minimum operating temperature of the equipment is limited to 5 10 C above ambien...

Page 17: ...th non flammable and flammable heat transfer liquids Alcohol and silicone oils are the most common flammable heat transfer liquids Use special care when using with these flammable liquids including th...

Page 18: ...mit Ht The temperature setting beyond the scope will not be accepted Ht This parameter is an absolute value and used for reminding only When temperature in the equipment is higher than the SEt value H...

Page 19: ...Auto Select between AoFF and Aon which means manually starting and automatically starting Atc Absolute temperature calibration ATC when the temperature displayed differs from the temperature measured...

Page 20: ...r may be accompanied by the following Ht When temperature in the bath is higher than the set value Ht is displayed Lt When temperature in the bath is lower than the set value Lt is displayed Er01 Warn...

Page 21: ...the digits or setting by pressing accept the setting by pressing or confirm and exit automatically in 4 seconds 5 Set the programming parameters Press and at the same time then release the buttons 1...

Page 22: ...en the equipment is in running mode timer for ASd is valid while the equipment is in standby mode timer for AS is valid The time is set by AS and ASd 8 Move between settings or increase the value Pres...

Page 23: ...ipment is in operation and start to heat When the set temperature and actual temperature are quite different heating LED will be steadily lit with full power heating and the temperature will go up wit...

Page 24: ...liquid The first time commissioned for use or when exchange of the heat transfer liquid and immediately after startup of the circulating pump check if any liquid leakage exists in the external circul...

Page 25: ...priate for calibration 3 5 Setting for restart mode after power failure Auto Press for more than 2 seconds until screen displays Ht Scroll the menu by until Auto is displayed then press to enter Use t...

Page 26: ...The function of temperature programming is optional 2 Press and at the same time and then release The parameter of first segment t1 is displayed Press to display original time press for a new value co...

Page 27: ...or disengaging the residual current circuit breaker RCD Pull the temperature of the heat transfer liquid as close as possible to the ambient before attempting to shut down the equipment except when o...

Page 28: ...onnect the power immediately Power on only after water or foreign material is completely evaporated or removed 4 2 Regular check of the STB Regular check of the STB is required for safe operation Inte...

Page 29: ...capacity as rated and do as follows 1 Make sure the mains plug has been removed 2 Loosen the two screws on the top and 4 screws at the back of the controller as shown in Fig 7 Remove the cove plate an...

Page 30: ...00 checked or replaced and for servicing or maintenance call authorized service engineer When the equipment displays Stb try first to pull down the temperature of the heater or the heat transfer liqui...

Page 31: ...and microprocessor based precise dynamic temperature control with temperature stability up to 0 05 C Independent immersion controller head DIY for precise temperature system with different bath capac...

Page 32: ...olute Temperature Calibration Auto start up or standby after power failure Duty cycle Continuous Wetted materials Stainless steel deep drawn Gastro norm pan Heater and pump of SUS316L Enclosure materi...

Page 33: ...DSGO0708 2 V211119 UXP8011 CD6400 CLASS I NFL RT 5 100 C...

Reviews: