5.2 Specifications
Models
UXP8011-HC090
UXP8011-HC130
UXP8011-HC250
Temperature range
°C (°F)
RT+10~+100 (RT+18~+212)
Temperature stability
±°C (°F)
0.02~0.05 (0.04~0.09)
Pump flow rate max L/min (gpm)
20 (5.3)
Pump pressure max
mbar (psi)
350 (5.1)
Temperature programming
6 segments×99h59min, 99 cycles
Classification for HTL to DIN 12876
Class 1/NFL, non-flammable Heat Transfer Liquid, No silicone oil
Safety
Protection and warning against short/open-circuit PT100
High-temperature cut off and low temperature alarm,
Protection against stalled pump motor & dry heating
Utility function
ATC (Absolute Temperature Calibration),
Auto start-up or standby after power failure
Duty cycle
Continuous
Wetted materials
Stainless steel deep drawn Gastro norm pan
Heater and pump of SUS316L
Enclosure material
SPCC mild steel with powder coating
EMC for USA and Canada
Class A, for operation only on networks without connected domestic areas
Heater wattage
W
1,000
Capacity
L (G)
6.5~9 (1.72~2.38)
9~13 (2.38~3.43)
18~25 (4.76~6.60)
Bath size (WDH)
mm (in)
150×280×200
(Gastro N Pan 1/3 - 8)
240×280×200
(Gastro N Pan 1/2 - 8)
280×480×200
(Gastro N Pan 1/1 - 8)
Immersion depth (H)
mm (in)
200 (8)
Bath opening (WD)
mm (in)
145×130
(5.7×5.2)
233×150
(9.2×5.9)
295×350
(11.6×13.8)
Overall size (WDH)
mm (in)
186×337×430
(7.3×13.3×16.9)
275×337×430
(10.8×13.3×16.9)
335×538×430
(13.2×21.2×16.9)
Packing size (WDH)
mm (in)
265×455×520
(10.4×17.9×20.5)
Net weight
Kg (lbs)
9 (19.8)
Gross weight
Kg (lbs)
11 (24.3)
Voltage and frequency
115V~, 60Hz
Current draw
A
12