Chapter 3 Controller Descriptions and Operation
3
Chapter 3 Controller Descriptions and Operation
3.1 Controller panel
Fig 6 Controller panel
60 Reset button (for resetting when alarming, the
circle indicating the location)
61 Display for temp and working parameters
62 Status indicators
63 Key board
64 Power switch
3.1.1 Display for temperature and working parameters
Normally it displays the equipment temperature once switched on and initiated or
oFF
when
standstill.
Other possible symbols include:
SEt
,
Ht
,
Lt
,
CooL
,
Auto
,
Atc
AS
,
ASd
,
cF
,
dEFt
,
ESc
,
Stb
.
Related definitions are as follows:
SEt
: Setting of working temperature, must be 5 °C higher than low limit
Lt
and 5 °C lower than
the high limit
Ht
. The temperature setting beyond the scope will not be accepted.
Ht
:
This parameter is an absolute value and used for reminding only. When temperature in the
equipment is higher than the
SEt
value,
Ht
is displayed. The controller will trigger a beeping
alarm and disable the heating. While the actual temperature is higher than the
Ht
setting, the
Ht
may be set to a higher value by pressing
so that
Ht
beeping can be cleared. If the
actual temperature is recovered to lower than the
Ht
setting, the
Ht
alarm will disappear
automatically with seconds of delay, or may be cleared immediately by pressing
. The
heating is possible only if the
Ht
is cleared.
Lt
:
This parameter is an absolute value and used for reminding only. When temperature in the
equipment is lower than the
SEt
value,
Lt
is displayed. The controller will trigger a beeping
alarm. While the actual temperature is lower than the
Lt
setting, the
Lt
may be set to a
lower value by pressing
so that
Lt
beeping can be cleared. If the actual temperature is