Chapter 2 Installing and Operating
2
2.2.2 Back view
Fig 2 Back view
03 Fresh-air inlets
04 Exhausting-air outlets
18 Handle
22 Power cable cord
23 Drain device
The inlets of fresh air and outlets of exhausting air of the equipment shall be kept clear of any
obstacles, so that no close-loop of the air circulating may develop.
If no external circulating is necessary, use hose, fitting and coupling to short-circuit the liquid
connections as for an external application system!
The equipment is not applicable for specimen or application system which develops tremendous
energy as result of reacting, dissolving or mixing under operating temperature range for
normal use! However for temperature below ambient until minimum temperature range of
the controller, optional cooling coil must be employed with additional external cooling source
available.