Chapter 1 Safety Warning and Instructions 

1

 

The maximum filling volume or cooling capacity of an open bath or refrigerated bath for 
movable immersion circulator shall be 25L and 600W. The fixing and connection for the 
immersion circulator, external circulating adaptor and the bath must be secured and 
hydraulically tight. Bath cover should be used where necessary to reduce the effect of 
evaporation of the liquid on the electronics of the circulator.  

Caution: smaller bath may develop hazard of overflowing, and leakage or slipping-off of the 
circulating tubes may cause hazard of spillage or spray! Excessive volume or capacity may 
degrade specifications for temperature control! 

Before attempting to remove the immersion circulator away from the bath, always unplug the 
equipment and wait until the temperature recovers close to ambient and note the following: 

1.

 

Don’t touch the heated surface or any parts with hot residual liquid or, put it on surface of 
flammable or temperature sensitive materials! Warning, Risk of burn hazard! 

2.

 

Don’t touch the frozen surface or any parts with cold residual liquid! Warning, Hazard of 
frostbite! 

Caution, Hazard of possible electric shock when removed and placed in horizontal or up-side 
down conditions and if the heat transfer liquid penetrates to electrical assembly resulting in 
deterioration of the insulation! 

 

Caution: Risk of frozen hazard! 

When water is employed as heat transfer liquid, and when intended for standstill for long 
period and when experiencing low ambient is possible, run the equipment at operating 
temperature close to ambient for at least 1hr. Empty the bath containing water, and remove 
the water completely. Disconnect the equipment and remove the mains plug from power 
source.  

 

Do not disassemble or exchange assembly parts and electric circuits by unauthorized 
personnel! Hazard of possible electric shock! In case of repairing, contact your local distributors 
or supporting technicians. 

Always remove the mains plug from the socket before cleaning, moving, maintenance or 
repairing! 

 

The first time powering up after repairing or standstill for long period, or at designated intervals 
under normal use, check the functionality of the following parts or components: 

Hydraulic tightness and fastening of drain device at three months intervals; 

Cutoff capability of the safety temperature limiter (STB) at 6 months intervals. 

Summary of Contents for UXP 8011-HC090

Page 1: ...SHANGHAI 200032 CHINA TEL 86 21 6443 1641 FAX 86 21 6403 9005 EMAIL BC XTEMP USA CN XTEMP ENVIRONMENTS SHANGHAI CO LTD ADD ROOMS 909 11 NO 2777 ZHOUHAI ROAD PUDONG NEW DISTRICT SHANGHAI 200137 CHINA T...

Page 2: ...it 14 2 2 5 Accessory cooling coil Optional 15 2 2 6 Put into operation 15 Chapter 3 Controller Descriptions and Operation 17 3 1 Controller panel 17 3 1 1 Display for temperature and working paramete...

Page 3: ...08 2 V211119 UXP8011 CD6400 CLASS I NFL RT 5 100 C Contents C 4 2 Regular check of the STB 27 4 3 Checking and replacing the fuse 28 4 4 Clear error warning and alarm 29 Chapter 5 Equipment Overview a...

Page 4: ...trical energy The operator is largely protected through the application of the appropriate standard specifications Further hazard sources may arise from the type of material being treated e g mixing d...

Page 5: ...he conditions that the equipment is operated and serviced exclusively in accordance with its intended purpose and as described in these Operating Instructions the equipment is not modified only origin...

Page 6: ...egrade electric safety of the equipment Note the effect of the altitude on the boiling point of the heat transfer liquids 1 2 Safety warning When heating and conditioning with specimen or application...

Page 7: ...clear of any obstacles Keep the equipment away from hot air emitting source direct sunshine strong magnetic field and electric sparks Neither contamination of high concentration dust or corrosive gas...

Page 8: ...iquids Use special care when using with these flammable liquids including the proper setting of the adjustable overheating protection device The circulation of the heat transfer liquid is important fo...

Page 9: ...kink and other mechanical stress develops during normal use which otherwise may result in thermal and or liquid hazards 2 Flexible hose fitting and coupling are important for the safe operation of the...

Page 10: ...azard 2 Don t touch the frozen surface or any parts with cold residual liquid Warning Hazard of frostbite Caution Hazard of possible electric shock when removed and placed in horizontal or up side dow...

Page 11: ...servicing or repairing Accessories lists UXP8011 HC090 Part Number Qty Descriptions Remarks DSGO0708 2 1 Operating instructions This document ERSG0202 1 Fuse 250V 15A D5L20 Breaking capacity 1500A BJG...

Page 12: ...f the equipment with protective ground PE For possible wet location conditions where cycle of evaporating and condensing overflowing or spillage may occur and for protection against hazard of possible...

Page 13: ...ment Use only hose fitting and coupling provided with the equipment or recommended by Xtemp The ratings of pressure and temperature for the hose fitting and the jacket container or heat exchanger of t...

Page 14: ...stacles so that no close loop of the air circulating may develop If no external circulating is necessary use hose fitting and coupling to short circuit the liquid connections as for an external applic...

Page 15: ...ne adaptor SL The connection to the pump outlet and external circulating kit shall be hydraulically tight Loose connection may result in hazard of liquid spray The pump inlet shall be clear of any obs...

Page 16: ...coil Optional Fig 5 Cooling coil For immersion heating bath GP series and circulators completed with open bath tank the minimum operating temperature of the equipment is limited to 5 10 C above ambien...

Page 17: ...th non flammable and flammable heat transfer liquids Alcohol and silicone oils are the most common flammable heat transfer liquids Use special care when using with these flammable liquids including th...

Page 18: ...mit Ht The temperature setting beyond the scope will not be accepted Ht This parameter is an absolute value and used for reminding only When temperature in the equipment is higher than the SEt value H...

Page 19: ...Auto Select between AoFF and Aon which means manually starting and automatically starting Atc Absolute temperature calibration ATC when the temperature displayed differs from the temperature measured...

Page 20: ...r may be accompanied by the following Ht When temperature in the bath is higher than the set value Ht is displayed Lt When temperature in the bath is lower than the set value Lt is displayed Er01 Warn...

Page 21: ...the digits or setting by pressing accept the setting by pressing or confirm and exit automatically in 4 seconds 5 Set the programming parameters Press and at the same time then release the buttons 1...

Page 22: ...en the equipment is in running mode timer for ASd is valid while the equipment is in standby mode timer for AS is valid The time is set by AS and ASd 8 Move between settings or increase the value Pres...

Page 23: ...ipment is in operation and start to heat When the set temperature and actual temperature are quite different heating LED will be steadily lit with full power heating and the temperature will go up wit...

Page 24: ...liquid The first time commissioned for use or when exchange of the heat transfer liquid and immediately after startup of the circulating pump check if any liquid leakage exists in the external circul...

Page 25: ...priate for calibration 3 5 Setting for restart mode after power failure Auto Press for more than 2 seconds until screen displays Ht Scroll the menu by until Auto is displayed then press to enter Use t...

Page 26: ...The function of temperature programming is optional 2 Press and at the same time and then release The parameter of first segment t1 is displayed Press to display original time press for a new value co...

Page 27: ...or disengaging the residual current circuit breaker RCD Pull the temperature of the heat transfer liquid as close as possible to the ambient before attempting to shut down the equipment except when o...

Page 28: ...onnect the power immediately Power on only after water or foreign material is completely evaporated or removed 4 2 Regular check of the STB Regular check of the STB is required for safe operation Inte...

Page 29: ...capacity as rated and do as follows 1 Make sure the mains plug has been removed 2 Loosen the two screws on the top and 4 screws at the back of the controller as shown in Fig 7 Remove the cove plate an...

Page 30: ...00 checked or replaced and for servicing or maintenance call authorized service engineer When the equipment displays Stb try first to pull down the temperature of the heater or the heat transfer liqui...

Page 31: ...and microprocessor based precise dynamic temperature control with temperature stability up to 0 05 C Independent immersion controller head DIY for precise temperature system with different bath capac...

Page 32: ...olute Temperature Calibration Auto start up or standby after power failure Duty cycle Continuous Wetted materials Stainless steel deep drawn Gastro norm pan Heater and pump of SUS316L Enclosure materi...

Page 33: ...DSGO0708 2 V211119 UXP8011 CD6400 CLASS I NFL RT 5 100 C...

Reviews: