31
STATus:QUEStionable:TEMPerature:ENABle <value>
This command sets the Questionable Status Temperature Enable Register to the specified value. The value is specified in
decimal form described in the description of “STATus:QUEStionable:TEMPerature:CONDition?”. A value 0 clears all Enables.
STATus:QUEStionable:TEMPerature:ENABle?
This command returns the Questionable Status Temperature Enable Register contents. The returned value is in decimal form
described in the description of “STATus:QUEStionable:TEMPerature:CONDition?”. If multiple bits are set, the result is the sum of
the set bit values.
STATus:QUEStionable:TEMPerature[:EVENt]?
This command returns the Questionable Status Temperature Event Register contents. The returned value is in decimal form
described in the description of “STATus:QUEStionable:TEMPerature:CONDition?”. If multiple bits are set, the result is the sum of
the set bit values.
SYSTem:AOUTput:MODE <DISabled|PARallel|SERies>
This command controls the function of the analog output of the power supply to facilitate series or parallel operation. In parallel
mode the analog output is proportional to the output current of the power supply, in series mode the analog output is
proportional to the output voltage of the power supply.
SYSTem:AOUTput:Mode?
This command returns the current state of the analog output of the power supply.
SYSTem:CAPability?
Display the unit capability string. For PLS600, the string (DCPSUPPLY WITH MEASURE) is returned.
SYSTem:CONFiguration:SAVE
This command saves configuration parameters to non-volatile memory. Contents consist of:
•
Analog current scaling select (3V/5V/10V)
•
Analog current calibration slopes
•
Analog voltage scaling select (3V/5V/10V)
•
Analog voltage calibration slopes
•
Auto-start (Enabled/Disabled)
•
Calibrated flag (Calibrated/Not Calibrated)
•
Control source (Local/Remote/Remote with Lock/Analog Voltage Control/Analog Current Control/Analog Dual Control)
•
Conversion current user slope and offset
•
Conversion current factory slope and offset
•
Conversion voltage user slope and offset
•
Conversion voltage factory slope and offset
•
Model
•
Over-current protection level
•
Over-power protection level
•
Over-voltage protection level
•
Power setpoint
•
Remote sense (Enabled/Disabled)
•
Remote sense lead resistance
•
Master mode
•
Unit serial number
•
LAN configuration
This command can be executed from any of the six operating modes (Local, Remote, Remote with Lock, Analog Voltage Control,
Analog Current Control, or Analog Dual Control), but only when the output is deactivated.
SYSTem:IFC:IPAddress? – This command returns the IP address of the power supply.