Error Messages
975-0200-01-01
B-5
B
Query Error List
An error number in the range [-499, -400] indicates that the output queue
control of the instrument has detected a problem with the message
exchange protocol described in IEEE 488.2, chapter 6. The occurrence of
any error in this class causes the query error bit (bit 2) in the Event Status
Register to be set.
-330
Self-test failed.
-350
Queue overflow.
A specific code entered into the queue in lieu of the code that
caused the error. This code indicates that there is no room in the
queue and an error occurred but was not recorded.
-360
Communication error.
This is the generic communication error for errors which
cannot be classified below.
-361
Parity error in program message.
Parity bit not correct when data received.
-362
Framing error in program message.
A stop bit was not detected when data was received, e.g. a data
rate mismatch.
-363
Input buffer overrun.
Software or hardware input buffer on serial port overflows with
data caused by improper (or nonexistent) pacing.
Table B-3
Device-Specific Error List
Error Code Error Message Description
Table B-4
Query Error List
Error Code Error Message Description
-400
Query Error
This is the generic query error for the power supply, used only
when the other types of errors do not apply.
-410
Query INTERRUPTED
Generated when a new command was received before it could
finish the query.
Summary of Contents for XTR6-110, XTR8-100, XTR12-70,
Page 2: ......
Page 3: ...XTR 850 Watt and 1700 Watt Series Programmable DC Power Supply Operating Manual...
Page 6: ......
Page 18: ...xvi...
Page 48: ...2 18...
Page 122: ...4 30...
Page 208: ...6 20...
Page 244: ...A 36...
Page 250: ...B 6...
Page 264: ...C 14...
Page 272: ...IX 4...
Page 273: ......