C-6
975-0200-01-01
Speci
fications
Remote Operation
Table C-3
Remote Operation
Programming Mode
APG
ISOL
Digital
Voltage and Current Output Voltage Programming
0–100%, 2~ up to 10 V, programmable
Current Output Resistor Programming
0–100%, 2~ up to 10 k
Ω
, programmable
Voltage Output Resistor Programming
0–100%, 2~ up to 10 k
Ω
, programmable
Output Voltage and Current Monitor
0–100%, 2~ up to 10 V, programmable
Voltage Programming Accuracy (mV)
± 0.5% of rated output voltage
±0.1% of rated output voltage
Current Programming Accuracy (mA)
± 1% of rated output current
±0.2% of rated output current
Voltage Readback Accuracy (mV)
± 1% of rated output voltage
±0.1% of rated output voltage
Current Readback Accuracy (mA)
± 1% of rated output current
±0.2% of rated output current
Isolation (Prog and Readback Lines)
With respect to chassis potential:
500 V
With respect to:
chassis potential: 500 V.
negative or positive main output: 1500 V.
negative or positive auxiliary output:
300 V.
Voltage and Current Programming Resolution
0.012% of full scale
Voltage and Current Readback Resolution
Parallel Operation
Up to 4 units in master/slave
mode
Series Operation
Up to 2 units (with external
diodes)
Constant Voltage (CV)/Constant Current (CC) Indicator
CV: TTL High (4–5 V)
CC: TTL Low (0–0.6 V)
Summary of Contents for XTR6-110, XTR8-100, XTR12-70,