Local Operation
3-18
975-0200-01-01
Clearing Alarms
Clearing Triggered and Manual Alarms
To clear a triggered alarm, use one of the following methods:
• Turn the power supply Off and then On.
• Press and hold the Rotary knob/Enter button for 3 seconds.
• Press the reset key combination (Flash Fail ONLY). See “Resetting the
Power Supply” on page 3–35.
To clear a manual alarm:
1. Press the Rotary knob/Enter button for 3 seconds.
The
Alr Clr
message will be displayed on the front panel for a few
seconds.
2. Repeat the alarm clearing until you have cleared all alarms.
Clearing a Flash Failure Alarm
The Flash Failure alarm is the only exception for clearing alarms. Flash
failure can only be cleared by a reset (see “Resetting the Power Supply”
on page 3–35) or by power cycling the unit.
To clear a triggered alarm, use one of the following methods:
•
Turn the power supply Off and then On.
•
Press and hold the Rotary knob/Enter button for 3 seconds.
•
Press the reset key combination (Flash Fail ONLY). See “Resetting
the Power Supply” on page 3–35.
The SCPI command (s) for these instructions are:
[:]OUTPut[<channel>]:PROTection:CLEar
Important:
When an alarm is cleared by using this method, only the current
highest precedence alarm is cleared. If another alarm has occurred, then this
alarm will not become the highest precedence alarm and will begin blinking on
the display.
Summary of Contents for XTR6-110, XTR8-100, XTR12-70,
Page 2: ......
Page 3: ...XTR 850 Watt and 1700 Watt Series Programmable DC Power Supply Operating Manual...
Page 6: ......
Page 18: ...xvi...
Page 48: ...2 18...
Page 122: ...4 30...
Page 208: ...6 20...
Page 244: ...A 36...
Page 250: ...B 6...
Page 264: ...C 14...
Page 272: ...IX 4...
Page 273: ......