23
5 Measuring the concentration
→
Call up the concentration measuring mode by
pressing .
→
Insert the round cell with barcode into the round
cell shaft until it clicks into place.
Align the line mark to the notch of the photome-
ter.
The photometer reads the barcode of the round cell
and automatically selects the relevant method.
Method and citation form are displayed in the header
of the display.
After the measurement is finished:
The measured value appears on the display.
If the select method menu appears, the
photometer has not yet recognized the
barcode of the AutoSelector. Align the
line mark of the AutoSelector to the notch
of the photometer.
5
M
eas
ur
in
g th
e c
onc
en
tr
a
tio
n
For photometrical concentration mea-
surements, it is always necessary to pre-
pare the sample to be measured.
Information on sample preparation can
be found in the first part, General infor-
mation and in the third part, Analytical
procedures of this operating manual.
Warning
It is essential to follow the safety instruc-
tions and notes on protective measures
when handling chemicals, in order to
avoid damage to your health and the en-
vironment. They can be found on the
packages of the coded cell tests and re-
agent test sets.
5.1
Measuring using coded cell tests
Line
mark
Barcode
Measured values outside the specified
measuring range are output in small nu-
merals.
Summary of Contents for PhotoLab Spektral
Page 2: ......
Page 32: ......
Page 38: ...8 List of contents...
Page 128: ...98 13 User defined methods...
Page 132: ...102 14 Photometer setup...
Page 140: ...110 16 RS 232 C interface...
Page 144: ...114 18 Exchanging the Memory Card...
Page 152: ...122 21 Appendix 21 1 Menu structure of the setup menu...
Page 153: ...123 21 Appendix...
Page 154: ...124 21 Appendix...
Page 160: ......
Page 161: ......