84
11 Correction functions
→
In the correction functions menu, call up the refer-
ence transmission menu item.
The following display appears:
The function, reference transmission is active and
appears in the special functions menu.
→
To measure the reference transmission, call up the
special functions submenu in the setup menu.
reference transmission function:
→
Select the menu item, On
→
Confirm with
.
→
Call up the reference transmission menu item.
→
Call up the measure ref. trm. menu item.
11.3
Reference transmission
Each transmission measurement is made against the
basic transmission stored in the meter.
When measuring the reference transmission, this val-
ue is ignored and the value measured as the reference
transmission is used instead.
The reference transmission function is not active
when delivered.
The measured reference transmission value remains
stored until
●
the photometer is switched off
●
the cell type (10 mm, 20 mm, 50 mm, round) is
changed
●
the wavelength is changed
●
the measuring mode, reference transmission, is
left
●
it is manually erased using the erase ref. transmis-
sion function
●
the function, reference transmission, is switched
off.
The erase ref. trm. menu item first appears
after at least one reference absorbance has
been measured.
Summary of Contents for PhotoLab Spektral
Page 2: ......
Page 32: ......
Page 38: ...8 List of contents...
Page 128: ...98 13 User defined methods...
Page 132: ...102 14 Photometer setup...
Page 140: ...110 16 RS 232 C interface...
Page 144: ...114 18 Exchanging the Memory Card...
Page 152: ...122 21 Appendix 21 1 Menu structure of the setup menu...
Page 153: ...123 21 Appendix...
Page 154: ...124 21 Appendix...
Page 160: ......
Page 161: ......