9 Analytical Quality Assurance (AQA)
The PipeCheck rapidly and easily determines any
potential pipette errors. While doing so, the photome-
ter identifies pipettes with volume errors of more than
5 % as faulty.
A Spectroquant®
PipeCheck is required to perform
the PipeCheck.
AQA entry
Using AQA MemoChip...
...or via a menu
→
In the setup menu, call up the AQA check sub-
menu.
→
Call up the PipeCheck submenu.
The following display appears:
→
Insert the AQA MemoChip.
→
Insert the reference cell.
After approx. 5 s:
9.2.4
Performing the PipeCheck