Operation
inoLab pH/ION 735
68
4.7
Transmitting data (RS 232 interface)
4.7.1
Options for data transmission
Via the RS 232 interface, you can transmit data to a PC or an external
printer. The following table shows which data are transmitted to the in-
terface in which way:
Note
The following rule applies: With the exception of the menus, shortly
pressing the
p
key generally outputs the display contents to the inter-
face (displayed measured values, measuring datasets, calibration
records).
Data
Control
Operation / description
Current
measured values
of the connected
sensor
Manual
!
By pressing
p
shortly.
!
Simultaneously with every
manual storage process
(see section 4.6.1).
Automatic, at
intervals
!
By pressing
p
for a long
time. Then you can set the
transmission interval.
!
Simultaneously with every
automatic storage process
(see section 4.6.2).
Stored measured
values
Manual
!
Displayed dataset with
p
after calling up from the
storage.
!
All datasets according to
the filter criteria via the
Out-
put to RS232
function.
For details, see section 4.6.3.
Calibration
records
Manual
!
Calibration record of a sen-
sor with
p
(after calling up
from the storage or at the
end of a calibration).
!
All calibration records after
calling up from the storage
via the
Output to RS232
function.
For details, see section 4.6.5.
Automatic
!
For the respective sensor
at the end of a calibration.