Operation
inoLab pH/ION 735
66
4.6.5
Displaying and outputting calibration records
You can display the calibration data
!
of one selected sensor, or
!
of all sensors connected
and then output them to the interface.
Displaying the calibra-
tion record of a selected
sensor
The calibration record of the last calibration can be found under the
Cal-
ibration
/
Calibration record
menu item in the respective measuring
menu. To open the settings, activate the relevant measuring window in
the measured value display and press the
g
key shortly.
Displaying the calibra-
tion records of all
sensors connected
The calibration records of the last calibration can be found under the
Data storage
/
Calibration data storage
menu item in the system menu.
To open the system menu in the measured value display, press the
g
key for approx. 1 s.
Menu item
Setting/func-
tion
Description
Data storage
/
Calibration data stor-
age
/
Display
-
Displays in pages all cali-
bration records of all sen-
sors.
Further options:
!
Scroll through the cali-
bration records with
82
.
!
Output the displayed
calibration record to the
interface with
p
.
!
Quit the display with
#
or
g
.
!
Switch directly to the
measured value display
with
m
.
Data storage
/
Calibration data stor-
age
/
Output to RS232
-
Outputs to the interface the
calibration records of all
sensors.