inoLab pH/ION 735
Technical data
83
7
Technical data
7.1
General data
Dimensions
approx. 172 x 80 x 37 mm
Weight
approx. 0.3 kg (without power pack)
Mechanical structure
Type of protection
IP 66
Electrical safety
Protective class
III
Test certificates
cETLus, CE
Ambient conditions
Storage
- 25 °C ... + 65 °C
Operation
-10 °C ... + 55 °C
Climatic class
2
Power
supply
Rechargeable batter-
ies
4 x 1.2 V nickel metal hydride (NiMH),
type AA
Operational life
up to 500 h with one battery charging
Power pack
(charging device)
FRIWO FW7555M/09, 15.1432.500-00
Friwo Part. No. 1883259
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA
Output: 9 V = / 1,5 A
Connection max. overvoltage category II
Primary plugs contained in the scope of de-
livery: Euro, US, UK and Australian.
Serial
interface
Connection of the cable AK 340/B or AK 325/S
Baud rate
adjustable:
1200, 2400, 4800, 9600, 19200 Baud
Type
RS232, data output
Data bits
8
Stop bits
2
Parity
None
Handshake
RTS/CTS
Cable length
Max. 15m
Guidelines
and norms used
EMC
EC guideline 89/336/EEC
EN 61326-1:1998
EN 61000-3-2 A14:2000
EN 61000-3-3:1995; FCC Class A
Instrument safety
E.C. guideline 73/23/EEC
EN 61010-1 A2:1995
Climatic class
VDI/VDE 3540
IP protection
EN 60529:1991