inoLab pH/ION 735
Operation
39
Menu item
Possible set-
ting
Description
Calibration record
-
Displays the calibration
record of the last calibra-
tion.
Man. temperature
-20 ... +130 °C
Entry of the manually de-
termined temperature. For
measurements without
temperature sensor only.
ISE setup
/
AutoRead criterion
low
medium
high
Selection of AutoRead cri-
teria (see section 4.5.2).
ISE setup
/
Unit
mg/l
mol/l
mg/kg
ppm
%
Selection with which unit
the result will be displayed.
Method
Std. add.
Std. add. ++
Std. sub.
Sample add.
Sample sub.
Blank add.
Selection of the available
measuring methods.
Ion type
Ag, Br, Ca, Cd,
Cl, CN, Cu, F,
I, K, Na, NO3,
Pb, S
Selection of the ion type to
be measured (only with
Unit
: mol/l)
Density
0.001 ... 9.999
g/ml or kg/l
Adjustable density of the
test sample (only with
Unit
: mg/kg, ppm, %)