Alarm Name
Alarm Description Behaviour in
Alarm Condition
Attention
required
Priority Range Default Notes
Open (TC1)
open circuit or
disconnected
reading (4
dashes on
display)
External Fill
Input S/cct
Short cct or polarity
reversal on Ext Fill
line.
Sim Fill and Gas
Bypass disabled
Prompt
High
N/A
15mins Fixed No
adjustment.
Unauthorised
Access
Lid opened without
PIN code entry
No change
Prompt
High
N/A
N/A
Alarm must be
re-set by
supervisor
Gas Bypass
Thermocouple
Fault
Thermocouple is
open circuit or
disconnected
Local gas bypass
will not function
Prompt
High
N/A
N/A
M507CE-I only
No LN2 (Gas
Bypass Timer)
Liquid not detected
at vent valve within
pre-set time limit.
No Fill until liquid
detected.
Continue venting
Prompt
High
0-
60min
30 mins M507CE-I only
Gas Bypass
fault -Valve
Open?
Liquid still detected
at vent valve after
pre-set time.
Vent valve stuck
open?
Immediate
High
0-
15min
5 mins M507CE-I only
General Alarm
>20 Intermittent
alarms in 24hr
period
No Change
Prompt
High
N/A
N/A
Alarms which
persist for <
Remote timer
setting could go
undetected
without this.
High
Temperature
(TC2)
TC2 Higher than
pre-set temperature
No change
Prompt
High
-80 to
-175°C
-130°C
Thermocouple
Open(TC2)
Thermocouple 2 is
open circuit or
disconnected
Bad temperature
reading (4
dashes on
display)
Prompt
High
N/A
N/A
Remote Alarm
Relay released
within pre-set time
of first alarm
No change
Prompt
High
0-
60min
30 mins Defaults to 5
mins max in
case of Ex High
Alarm.