Instruction Manual
K SERIES Cryostorage System
IM: CryoCeK
Revision: 02
Approved by: CB
Page
38
of
87
No
Steps of procedure
Display and notes
3
Enter the four digit supervisor's code using
the numbered keys and the
and
keys
to move the cursor
Press
key to finish when the cursor is on
the last digit
If a code is requested and it is not known,
the password will have to be reset as per
section 7.3
If the code is accepted (or not requested) the
following menu will appear:
NOTE On initial installation setup, first configurate the required features in the
CONFIG/CALIBRATE
menu, and then go back to the
TIMERS/CLOCK
menu!
4
Press switch 2 to
CONFIGURATE/CALIBRATE
If the CONFIG/CALIBRATE menu is selected
the following message (example) will appear:
5
Use the + and
– keys to select the
LANGUAGE:
English
German
French
Press the
key when finished
If the calibration of the first thermocouple
is disabled skip to step 10
If the calibration is enabled (see Engineering
menu as per section 7.4) the following
message will appear:
NOTE Thermocouples are calibrated using melting ice as a 0 °C reference and liquid
nitrogen to set the gain. Ensure both are to hand and within reach of the thermocouple tip
before proceeding.
6
Press switch 1 to select CALIBRATE TC1
To skip over calibration press the
key
and go to step 10
The following message will appear:
NOTE Place the thermocouple in the ice and allow a few seconds for the temperature to
stabilise before continuing ice calibration.