Instruction Manual
K SERIES Cryostorage System
IM: CryoCeK
Revision: 02
Approved by: CB
Page
30
of
87
Table 6
Procedure for setting up the controller
No
Steps of procedure
Display and notes
1
Press
key to enter the menu
The Firmware Version number and serial
number will be briefly displayed followed by the
menu:
2
Press switch 2 to CHANGE SETTINGS
If PASSWORD PROTECTION is disabled
go to step 4
If password protection is enabled the following
message will appear:
3
Enter the four digit supervisor's code using
the numbered keys and the
and
keys
to move the cursor
Press
key to finish when the cursor is on
the last digit
If a code is requested and it is not known,
the password will have to be reset as per
section 7.3
If the code is accepted (or not requested) the
following menu will appear:
NOTE On initial installation setup, first configurate the required features in the
CONFIG/CALIBRATE
menu, and then go back to the
TIMERS/CLOCK
menu!
4
Press switch 1 to set up timers and/or
clock
If the TIMERS/CLOCK menu is selected the
following message (example) will appear:
If the temperature of the thermocouple rises above the displayed value an alarm will be
activated.
5
Use the + and - keys to set the alarm
temperature:
the ALARM TEMP setting is adjustable
between 0 °C and
–175 °C in 5-degree
If the Temperature Control Option is enabled
(see CONFIG/CALIBRATE menu) the following
message (example) will appear: