Instruction Manual
K SERIES Cryostorage System
IM: CryoCeK
Revision: 02
Approved by: CB
Page
40
of
87
No
Steps of procedure
Display and notes
followed by:
9
The calibration is complete
Press switch 1 to repeat the calibration; or
Press the
key to proceed to next step; or
Press the
key to exit
NOTE It is recommended that the calibration is confirmed by checking the temperature
display with the thermocouple in ice and liquid nitrogen
.
10
If the temperature of the thermocouple is not
stable during the ice or LN2 calibration the
following message may appear:
If the temperature of the thermocouple is out of
an acceptable range (e.g. if it is not correctly
inserted into the ice or LN2 during calibration)
one of the following message (examples) may
appear:
Following any of the above messages will
return to step 5, i.e.: