Instruction Manual
K SERIES Cryostorage System
IM: CryoCeK
Revision: 02
Approved by: CB
Page
52
of
87
7.7 CONTROLLER RESET
The RESTARTING function can be used to force the controller to go through its start-up
sequence. This can be used for resetting the controller and clearing the data log.
NOTE
Resetting the controller clears the data log and all data will be lost.
Table 10
Procedure for resetting the controller
No
Steps of procedure
Display and notes
1
Make sure the PC cable (M519CE or
M549CE) is not plugged in, if used
Press the
key to enter the menu
After a few seconds the following message will
appear:
2
Press and hold the
key until the
controller will bleep three times
The following message will appear:
3
Press the
key to proceed or
Press the
↑
key to return to the
Temperature/Level display
If PASSWORD PROTECTION is enabled the
following message will appear:
4
Enter the four digit supervisor's code using
the numbered keys and the
and
keys
to move the cursor
Press
key to finish
If a code is requested and it is not known,
the password will have to be reset (see
section 7.3)
If the code is accepted (or not requested) the
following message will appear:
5
Use the
and
keys to select:
Yes or No
If No is selected the data memory is left
intact and the controller returns to the
Temperature/Level display
If Yes is selected the following message will
appear: