7. SYSTEM INSTALLATION
5)
Follow the schema bellow for the best measurement results. Make sure the black body and the camera are in
a stable environment with surrounding room temperatures between 15°C to 35°C. There can be no air circula on
in the near vicinity of the black body to ensure the homogenity of the black body heated surface.
BEST ACCURACY
SIDE VIEW
FRONT VIEW
ALTERNATIVE LOCATIONS
OF BLACK BODY
ALTERNATIVE LOCATIONS
OF BLACK BODY
FOV = 45° × 37°
Setting focus distance:
Distance from camera to black body depends on focus distance
(distance is equally like focus distance)
Focus distance A = 4 m (13.1 ft)
Focus distance B = 8 m (26.2 ft)
For focus distance A is 3–6 m (9.8–19.7 ft)
For focus distance B is 7–11 m (23–36.1 ft)
Recommended minimum
distance from the ground
1.6 m (5.2 ft)
MEASURING ZONE
DIST
ANCE FR
OM
THE GR
OUND
BLACK BODY LOCATION
for 4 m
(13.1 ft)
for 8 m
(26.2 ft)
Figure 7.3 – Workswell MEDICAS measurement scheme.
7.2 Power Supply
The Workswell MEDICAS system can be powered through a +9 to +36 VDC connector on the back panel. The
back panel DC connector is compa ble with a 5.5 mm/OD (outer diameter) and 2.5 mm/ID (inner diameter)
plug, where the inner contact is +9 to +36 VDC and the shell is GND.