9. Specifications
9. Specifications
Model CPH6300 hand-held pressure indicator (complete measuring chain)
Measuring inputs
1 input for CPH6300-S1
2 inputs for CPH6300-S2
Measuring range
Gauge pressure
mbar
-600 ... 0
-600 ... +600 -400 ... 0
-400 ... +400
-250 ... 0
-250 ... +250 -100 ... +100 -20 ... 60
-20 ... 40
-20 ... 25
0 ... 25
0 ... 40
0 ... 60
0 ... 100
0 ... 160
0 ... 250
0 ... 400
0 ... 600
bar
-1 ... 0
-1 ... 1,5
-1 ... 3
-1 ... 5
-1 ... 9
-1 ... 15
-1 ... 24
-1 ... 39
0 ... 1
0 ... 1.6
0 ... 2.5
0 ... 4
0 ... 6
0 ... 10
0 ... 16
0 ... 25
0 ... 40
0 ... 60
0 ... 70
0 ... 100
0 ... 160
0 ... 250
0 ... 400
0 ... 600
0 ... 1,000
psi
0 ... 0.4
0 ... 0.6
0 ... 0.9
0 ... 1.5
0 ... 2.5
0 ... 4
0 ... 6
0 ... 10
0 ... 14.5
0 ... 25
0 ... 40
0 ... 60
0 ... 90
0 ... 145
0 ... 250
0 ... 360
0 ... 580
0 ... 870
0 ... 1,450
0 ... 2,320
0 ... 3,630
0 ... 5,800
0 ... 8,700
0 ... 14,500
Absolute pressure
mbar abs. 0 ... 250
0 ... 400
0 ... 600
bar abs.
0 ... 1
0 ... 1.6
0 ... 2.5
0 ... 4
0 ... 6
0 ... 10
0 ... 16
0 ... 25
0.8 ... 1.2
Overpressure limit
(depending on
measuring range)
3 times; ≤ 25 bar
2 times; > 25 bar ... ≤ 600 bar
1.5 times; > 600 bar
3 times; ≤ 360 psi
2 times; > 360 psi ... ≤ 8,700 psi
1.5 times; > 8,700 psi
Resolution
dependant on pressure range (max. 4 1/2-digit)
Accuracy of the
measuring chain
1)
0.2 % FS (resolution 4-digit);
{optional: 0.1 % FS (resolution 4 1/2-digit)}
2)
{} Items in curved brackets are optional extras for an additional price.
1) It is defined by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following
factors: the intrinsic performance of the instrument, the measurement uncertainty of the reference instrument, long-term
stability, influence of ambient conditions, drift and temperature effects over the compensated range during a periodic zero point
adjustment.
2) Reference conditions: 15 ... 25 °C (59 ... 77 °F)