22

WIKA operating instructions, model CPH6300

EN

14043054.02 08/2017 EN/DE

As soon as a sufficiently long set time (in seconds) has been measured in order 

to calculate the mean value, 

____

 is shown in the display, in the lower display, a 

'countdown' is displayed.

During the low-power logger operation, the averaging function is always deactivated.

Function of Min/Max value memory in conjunction with the averaging function:

 

If averaging is activated and slow measuring, 

rAtE SLo

, is selected, the min/max 

memory value relates to the average display value.

 

If averaging is activated and fast measuring function is selected (

rAtE FASt

 

or 

P

.

dEt

), the min/max memory value relates to the internal measured values 

(> 1,000 Hz measuring frequency).

5.4.5  Rate of change of pressure (only for sensor connection 1)

With the 

CALC

 function 

dPdt

, pressure changes per time can be directly displayed. The 

underlying measurement intervals, 

t.int

, can thus be chosen freely. The instrument 

automatically saves pressure values over this time period and calculates, using the 

current measured value and the historical data in the memory, the rate of change of the 

pressure (pressure change/time).

[TARE] button:

 The historical data is deleted, calculation begins again from the start.

Example:

Leak testing for wastewater channels with the criteria of lower than -3 mbar/min over a 

test duration of 5 minutes with a test pressure of 250 mbar.

Equipment/default setting:

 

Shut-off bladder and pressure pumps

 

Sensor: CPT6200 (measuring range 0 … 1 bar)

 

 

bASE

 = 

PEt

.

n

 (minutes) Time unit for the pressure change display

 

 

t

.

int

 = 1:00 (1 minute) test duration, for which the pressure range display is 

calculated

 

Logger: 

CyCL

, 5 seconds

Execution:

1.  Connect components, make necessary sealing.

2.  Start logger.

3.  Load with pressure, wait sufficient time.

4.  Re-start calculation with 

[TARE]

 button.

 

Current pressure value is displayed.

5. Commissioning, operation

Summary of Contents for CPH6300-S1

Page 1: ...Operating instructions Betriebsanleitung EN DE Hand held pressure indicator model CPH6300 Hand Held Druckmessger t Typ CPH6300 Hand held pressure indicator model CPH6300...

Page 2: ...can be found at www wika com 08 2017 WIKA Alexander Wiegand SE Co KG All rights reserved Alle Rechte vorbehalten WIKA is a registered trademark in various countries WIKA ist eine gesch tzte Marke in v...

Page 3: ...5 1 Operation 11 5 1 1 Display 11 5 1 2 Controls 12 5 2 Commissioning 13 5 3 Configuration of the instrument 13 5 4 Special functions 20 5 4 1 Different pressure units 20 5 4 2 Different measuring mod...

Page 4: ...ntenance cleaning and recalibration 33 7 1 Maintenance 33 7 2 Cleaning 35 7 3 Recalibration 35 8 Dismounting return and disposal 36 8 1 Dismounting 36 8 2 Return 37 8 3 Disposal 37 9 Specifications 38...

Page 5: ...strument s range of use The operating instructions are part of the product and must be kept in the immediate vicinity of the instrument and readily accessible to skilled personnel at any time Pass the...

Page 6: ...l 2 4 Interface connector 5 Display 6 Controls 7 Reference pressure sensor model CPT6200 2 2 Description The model CPH6300 hand held pressure indicator is very well suited for operation under the most...

Page 7: ...uires high accuracy pressure measurement This instrument is not permitted to be used in hazardous areas The instrument has been designed and built solely for the intended use described here and may on...

Page 8: ...th abrasive or viscous media Observe the working conditions in accordance with chapter 9 Specifications Always operate the instrument within its overload limits Any use beyond or different to the inte...

Page 9: ...oduct label for model CPH6300 S1 or CPH6300 S2 examble Product label for model CPT6200 examble 1 Product name 2 Date of manufacture month year 3 Serial number and article number 4 Pressure measuring r...

Page 10: ...warm environment the formation of condensation may result in instrument malfunction Before putting it back into operation wait for the instrument temperature and the room temperature to equalise 4 2 P...

Page 11: ...ing 5 Commissioning operation 5 Commissioning operation Personnel Skilled personnel Only use original parts see chapter 10 Accessories 5 1 Operation 5 1 1 Display 1 Main display Current measured value...

Page 12: ...off MIN function on off 2 sec Delete MIN value Change the secondary display CH1 CH2 DIF only for 2 channel 2 sec Main menu Enter configuration if Hold function on off A Store measured value B 2 sec Cl...

Page 13: ...data 1 To configure press the SET MENU button for 2 seconds The menu main display SEt is called up 2 With the SET MENU button select the required menu branch 3 Select the parameter with the button 4...

Page 14: ...f on Sea level correction on see chapter 5 4 3 Sea level correction for absolute pressure sensors ALT 2000 9999 Sea level correction in m only when sensor 1 SL on rate Rate Measuring rate 1 Slo Slow S...

Page 15: ...on function oFF No calculated display dP dt Rate of change of pressure SPEd Airspeed through the orifice plate pitot tube FLo Flow Air flow dP dt Settings base Time unit for the rate of change of the...

Page 16: ...000 0 00 00 Decimal point setting strt oFF 1 1000 Minimum display below which display 0 is forced SET Out Set output Settings for universal output Out oFF Interface and analogue output deactivated SE...

Page 17: ...or 2 000 2 000 The measuring scale will be changed by this factor to compensate deviations of probe measuring instrument SET Corr Set Corr Adjustment of the measurement 2 channel version 1 3 OFS I Zer...

Page 18: ...ax Max alarm limit sensor 1 not if AL 1 oFF Sensor1 Max is the upper scale range limit of sensor 1 AL 2 On No So Alarm sensor 2 on with buzzer without buzzer OFF No alarm function for sensor 2 AL Lo 2...

Page 19: ...e logger oFF No logger function CYCL 00 01 60 00 Cycle time for cyclic logger minutes seconds 1 Lo Po on oFF Low power logger with very low power supply only if cyclic logger with slow measuring rate...

Page 20: ...3 different measuring modes for various purposes Two of these P dEt and FASt operate with an increased measuring frequency of 1 000 measurements s 5 4 2 1 Standard measurement slow rate sLo Measuring...

Page 21: ...ion see chapter 5 3 Configuration of the instrument The setting is only possible if an abs pressure sensor is connected to sensor connection 1 2 Input the height above sea level in metres see chapter...

Page 22: ...ime can be directly displayed The underlying measurement intervals t int can thus be chosen freely The instrument automatically saves pressure values over this time period and calculates using the cur...

Page 23: ...e display value can easily be misinterpreted since the display values react correspondingly slowly The display value is calculated P Tcurrent P Tcurrent t int t int If the data is not yet available ov...

Page 24: ...es are available at www wika com With the logger function Fvnc Strt or Fvnc CYCL activated the hold function is not available the STORE QUIT button is then responsible for the logger operation 5 5 1 M...

Page 25: ...y is full 5 5 1 2 Requesting manual recording Saved data sets can be read with GSOFT software and also be viewed in the instrument display With the single value logger after saving a measured value th...

Page 26: ...n of the instrument following the start of the logger the measured value is automatically at the interval of the set cycle time The logger cycle time is adjustable from 1 s 60 min At the moment of sav...

Page 27: ...nsor connection 1 Sensor 1 Arithmetic mean Sensor 1 Min Peak Max Peak since last save Sensor 2 Arithmetic mean Sensor 2 Min Peak Max Peak since last save Difference sensor 1 sensor 2 Arithmetic mean C...

Page 28: ...he process SToP yes Stop recording 3 Press STORE QUIT button Confirmation of the selection logger recording ended If you try to switch off an instrument during a logging cycle you will automatically b...

Page 29: ...t can be connected directly to a USB interface of a PC The transmission is made in a binary coded format and is protected against transmission errors by an extensive safety mechanism CRC As operation...

Page 30: ...L 100 5 8 Pressure connection at the pressure sensors The instrument has been designed so that all model CPT6200 sensors can be connected without the need for any recalibration This means that a large...

Page 31: ...taken out of operation immediately Ensure that pressure or signal is no longer present and protect against accidental commissioning Contact the manufacturer If a return is needed please follow the ins...

Page 32: ...equipment For contact details see chapter 1 General information or the back page of the operating instructions Display Meaning Remedy SEnS Erro or Err 9 No sensor connected Switch instrument off and c...

Page 33: ...not be calculated Choose different unit Calculation range exceeded Choose different unit Err 7 System error Send in for repair Sensor not present recognised Plug in inserted sensor once more with runn...

Page 34: ...etration Avoid unscrewing the instrument unnecessarily Carry out the battery change as described below 1 Unscrew the three cross head screws on the rear of the instrument 2 Lie the instrument still cl...

Page 35: ...o cleaning isolate the instrument properly from the pressure source and switch it off 2 Use the requisite protective equipment 3 Clean the instrument with a moist cloth Electrical connections must not...

Page 36: ...ies and damage to property and the environment through residual media Upon contact with hazardous media e g oxygen acetylene flammable or toxic substances harmful media e g corrosive toxic carcinogeni...

Page 37: ...trument see chapter 7 2 Cleaning When returning the instrument use the original packaging or a suitable transport packaging To avoid damage 1 Wrap the instrument in an antistatic plastic film 2 Place...

Page 38: ...ar abs 0 250 0 400 0 600 bar abs 0 1 0 1 6 0 2 5 0 4 0 6 0 10 0 16 0 25 0 8 1 2 Overpressure limit depending on measuring range 3 times 25 bar 2 times 25 bar 600 bar 1 5 times 600 bar 3 times 360 psi...

Page 39: ...the measuring range freely selectable Functions Measuring rate 4 s slow 1 000 s fast 1 000 s unfiltered peak value detection selectable Memory Min Max integrated data logger Functions via button press...

Page 40: ...odel CPT6200 Pressure connection 4 G B flush G 1 for 0 1 1 6 bar 1 5 25 psi or various connection adapters on request Material Wetted parts Stainless steel or Elgiloy 25 bar 360 psi additionally with...

Page 41: ...ainty of the reference instrument long term stability influence of ambient conditions drift and temperature effects over the compensated range during a periodic zero point adjustment 2 Reference condi...

Page 42: ...42 WIKA operating instructions model CPH6300 EN 14043054 02 08 2017 EN DE Dimensions in mm in Digital indicator CPH6300 S1 42 1 65 163 6 42 86 3 39 9 Specifications...

Page 43: ...43 WIKA operating instructions model CPH6300 EN 14043054 02 08 2017 EN DE Digital indicator CPH6300 S2 42 1 65 163 6 42 86 3 39 9 Specifications...

Page 44: ...or 2 Connection Channel 1 only with CPH6300 S1 3 Connection Channel 2 only with CPH6300 S2 4 Connection Channel 1 only with CPH6300 S2 Reference pressure sensor CPT6200 Model CPH6300 S1 Model CPH6300...

Page 45: ...nsor connection cable approx 1 1 m 3 3 ft CPH A 63 ZSZZZZZZZZ Z Extension cable for connection of sensors approx 3 8 m 12 5 ft to approx 5 m 16 4 ft CPH A 63 ZVZZZZZZZZ Z 2 connection cable 2 m 6 6 ft...

Page 46: ...46 WIKA operating instructions model CPH6300 EN 14043054 02 08 2017 EN DE...

Page 47: ...etrieb 55 5 1 Bedienung 55 5 1 1 Anzeige 55 5 1 2 Bedienelemente 56 5 2 Inbetriebnahme 57 5 3 Konfiguration des Ger tes 57 5 4 Sonderfunktionen 64 5 4 1 Verschiedene Druckeinheiten 64 5 4 2 Verschiede...

Page 48: ...ung Reinigung und Rekalibrierung 77 7 1 Wartung 77 7 2 Reinigung 79 7 3 Rekalibrierung 79 8 Demontage R cksendung und Entsorgung 80 8 1 Demontage 80 8 2 R cksendung 81 8 3 Entsorgung 81 9 Technische D...

Page 49: ...und allgemeinen Sicherheitsbestimmungen einhalten Die Betriebsanleitung ist Produktbestandteil und muss in unmittelbarer N he des Ger tes f r das Fachpersonal jederzeit zug nglich aufbewahrt werden Be...

Page 50: ...tstellenanschluss 5 Anzeige 6 Bedienelemente 7 Referenz Drucksensor Typ CPT6200 2 2 Beschreibung Das Hand Held Druckmessger t Typ CPH6300 ist durch sein robustes und wasserdich tes Design sehr gut zum...

Page 51: ...ung bei der eine hochgenaue Druckmessung erforderlich ist verwendet werden Dieses Ger t ist nicht f r den Einsatz in explosionsgef hrdeten Bereichen zugelassen Das Ger t ist ausschlie lich f r den hie...

Page 52: ...e Messstoffe verwenden Betriebsparameter gem Kapitel 9 Technische Daten beachten Ger t immer innerhalb des berlastgrenzbereiches betreiben Jede ber die bestimmungsgem e Verwendung hinausgehende oder a...

Page 53: ...kennzeichnungen Typenschild f r Typ CPH6300 S1 oder CPH6300 S2 Beispiel Typenschild f r Typ CPT6200 Beispiel 1 Produktname 2 Herstellungsdatum Monat Jahr 3 Seriennummer und Artikelnummer 4 Druckmessbe...

Page 54: ...sowie innerbetrieblichem Transport vorsichtig vorgehen und die Symbole auf der Verpackung beachten Bei innerbetrieblichem Transport die Hinweise unter Kapitel 4 2 Verpa ckung und Lagerung beachten Wir...

Page 55: ...e einen Beutel mit Trocknungsmittel der Verpackung beilegen 5 Inbetriebnahme Betrieb Personal Fachpersonal Nur Originalteile verwenden siehe Kapitel 10 Zubeh r 5 1 Bedienung 5 1 1 Anzeige 1 Hauptanzei...

Page 56: ...aus 2 Sek MIN Wert L schen Umschalten der Nebenanzeige CH1 CH2 DIF nur bei 2 Kanal 2 Sek Hauptmen Aufruf der Konfiguration wenn Hold Funktion ein aus A Messwert Speichern B 2 Sek Speicher Freigabe 2...

Page 57: ...ten enth lt 1 Zum Konfigurieren Taste SET MENU 2 Sekunden lang dr cken Das Men Hauptanzeige SEt wird aufgerufen 2 Mit Taste SET MENU gew nschten Men zweig w hlen 3 Mit Taste Parameter w hlen 4 Mit den...

Page 58: ...on Meeresh hen Korrektur an siehe Kapitel 5 4 3 H henkorrektur bei Absolut druck Sensoren ALT 2000 9999 Meeresh hen Korrektur in m nur wenn Sensor 1 SL on rate Rate Messgeschwindigkeit 1 Slo Slow lan...

Page 59: ...e berechnete Anzeige dP dt Druck nderungsgeschwindigkeit SPEd Luftgeschwindigkeit ber Blende Staurohr FLo Durchfluss Volumenstrom dP dt Einstellungen base Zeiteinheit der Druck nderungsgeschwindigkeit...

Page 60: ...alpunkt Einstellung strt oFF 1 1000 Mindestanzeige darunter wird Anzeige 0 erzwun gen SET Out Set Output Einstellungen f r universellen Ausgang Out oFF Schnittstelle und Analogausgang deaktiviert SEr...

Page 61: ...en 2 Kanal Version 1 3 OFS I Nullpunktkorrektur Offset des Sensors 1 oFF Keine Nullpunktkorrektur des Sensors 1 sensorabh ngig z B 5 00 5 00 mbar Der Nullpunkt des Sensors 1 wird um diesen Wert versch...

Page 62: ...sor 2 nicht wenn AL 2 oFF Sensor1 Min ist die untere Anzeigebe reichsgrenze des Sensors 2 AL Hi 2 AL 2 Lo Sensor2 Max Max Alarm Grenze Sensor 2 nicht wenn AL 2 oFF Sensor1 Max ist die obere Anzeigebe...

Page 63: ...ringer Stromaufnahme nur bei zyklischem Logger und langsamer Messung 1 SET CLOC Set Clock Einstellen der Echtzeituhr CLOC HH MM Clock Einstellen der Uhrzeit Stunden Minuten YEAr YYYY Year Einstellen d...

Page 64: ...as Ger t unterst tzt 3 verschiedene Messarten f r verschiedene Anwendungszwe cke Zwei davon P dEt und FASt arbeiten mit einer erh hten Messfrequenz von 1 000 Messungen s 5 4 2 1 Standardmessung slow r...

Page 65: ...tion des Ger tes Die Einstellung ist nur m glich wenn ein Absolutdrucksensor an Sensorbuchse 1 angeschlossen ist 2 Die H he ber dem Meeresspiegel in Meter eingeben siehe Kapitel 5 3 Konfiguration des...

Page 66: ...nktion dPdt k nnen Druck nderungen pro Zeit direkt angezeigt werden Die zugrundeliegenden Messintervalle t int k nnen dabei frei gew hlt werden Das Ger t speichert automatisch Druckwerte ber diesen Ze...

Page 67: ...k nnen Anzei gewerte leicht missverstanden werden da der angezeigte Wert entspre chend langsam reagiert Der Anzeigewert wird berechnet P Taktuell P Taktuell t int t int Wenn noch nicht Daten ber den...

Page 68: ...ka de verf gbar Bei aktivierter Loggerfunktion Fvnc Strt oder Fvnc CYCL steht die Hold Funktion nicht zur Verf gung die Taste STORE QUIT ist dann f r die Loggerbedienung zust n dig 5 5 1 Manuelle Aufz...

Page 69: ...gerspeicher voll ist 5 5 1 2 Manuelle Aufzeichnung abrufen Abgespeicherte Datens tze k nnen sowohl mit der Software GSOFT ausgelesen als auch in der Ger teanzeige selbst betrachtet werden Beim Einzelw...

Page 70: ...gew hlt siehe 5 3 Konfiguration des Ger tes werden nach Start des Loggers automatisch Messwerte im Abstand der eingestellten Zykluszeit aufgezeichnet Die Logger Zykluszeit ist einstellbar von 1 s 60 m...

Page 71: ...luss 1 Sensor 1 Arithmetischer Mittelwert Sensor 1 Min Peak Max Peak seit dem letzten Speichern Sensor 2 Arithmetischer Mittelwert Sensor 2 Min Peak Max Peak seit dem letzten Speichern Differenz Senso...

Page 72: ...yes Aufzeichnung stoppen 3 Taste STORE QUIT dr cken Best tigung der Auswahl Loggeraufzeichnung beendet Wird versucht ein mit zyklischer Aufzeichnung laufendes Ger t auszuschal ten wird automatisch na...

Page 73: ...ine USB Schnittstelle eines PC angeschlossen werden Die bertra gung erfolgt in einem bin r codierten Format und ist durch aufwendige Sicherheitsme chanismen gegen bertragungsfehler gesch tzt CRC Als B...

Page 74: ...ektur Anzeige gemessener Wert OFFS 1 SCAL 100 5 8 Druckanschluss an die Drucksensoren Das Messger t ist so konzipiert dass alle Sensoren des Typs CPT6200 ohne Neuab gleich angesteckt werden k nnen Som...

Page 75: ...rd immer mit Hupe gemeldet 5 Inbetriebnahme Betrieb 6 St rungen 6 St rungen Personal Fachpersonal Schutzausr stung Schutzhandschuhe und Schutzbrille Werkzeuge Schraubenschl ssel oder Drehmomentschl ss...

Page 76: ...ite der Betriebsan leitung Anzeige Bedeutung Abhilfe SEnS Erro oder Err 9 Es ist kein Sensor angesteckt Ger t ausschalten und Sensor anstecken Angesteckter Sensor oder Ger t ist defekt Mit evtl vorhan...

Page 77: ...nicht vorhanden erkannt Abgesteckten Sensor wieder anste cken bei laufender Aufzeichnung Logger stoppen und erneut starten Anzeigewert nicht berechenbar Passende Sensorkombination anste cken 6 St run...

Page 78: ...Kreuzschlitzschrauben an der R ckseite des Ger tes herausschrauben 2 Das noch geschlossenes Ger t so ablegen dass Anzeige sichtbar bleibt W hrend des gesamten Batteriewechsels das Ger t so liegen lass...

Page 79: ...tung verwenden 3 Das Ger t mit einem feuchten Tuch reinigen Elektrische Anschl sse nicht mit Feuchtigkeit in Ber hrung bringen VORSICHT Besch digung des Ger tes Eine unsachgem e Reinigung f hrt zur Be...

Page 80: ...NUNG K rperverletzungen Sach und Umweltsch den durch Messstoffreste Bei Kontakt mit gef hrlichen Messstoffen z B Sauerstoff Acetylen brenn baren oder giftigen Stoffen gesundheitsgef hrdenden Messstoff...

Page 81: ...tverpackung verwenden Um Sch den zu vermeiden 1 Das Ger t in eine antistatische Plastikfolie einh llen 2 Das Ger t mit dem D mmmaterial in der Verpackung platzieren Zu allen Seiten der Transportverpac...

Page 82: ...solutdruck mbar abs 0 250 0 400 0 600 bar abs 0 1 0 1 6 0 2 5 0 4 0 6 0 10 0 16 0 25 0 8 1 2 berlast Druckgrenze 3 fach 25 bar 2 fach 25 bar 600 bar 1 5 fach 600 bar 3 fach 360 psi 2 fach 360 psi 8 70...

Page 83: ...ei w hlbar Funktionen Messrate 4 s slow 1 000 s fast 1 000 s ungefiltert Spitzenwerterfas sung ausw hlbar Speicher Min Max integrierter Datenlogger Funktionen via Tastendruck Hintergrundbeleuchtung Mi...

Page 84: ...ntb ndig G 1 f r 0 1 1 6 bar 1 5 23 2 psi oder diverse Anschlussadapter auf Anfrage Werkstoff Messstoffber hrte Teile CrNi Stahl oder Elgiloy 25 bar 360 psi zus tzlich mit Dichtung NBR 3 Frontb ndige...

Page 85: ...sgedr ckt wird und fol gende Faktoren beinhaltet die ger tespezifische Performance Messunsicherheit des Referenzger tes Langzeitstabilit t Einfluss durch Umgebungsbedingungen Drift und Temperatureinfl...

Page 86: ...86 WIKA Betriebsanleitung Typ CPH6300 DE 14043054 02 08 2017 EN DE Abmessungen in mm in Digitales Anzeigeger t CPH6300 S1 42 1 65 163 6 42 86 3 39 9 Technische Daten...

Page 87: ...87 WIKA Betriebsanleitung Typ CPH6300 DE 14043054 02 08 2017 EN DE Digitales Anzeigeger t CPH6300 S2 42 1 65 163 6 42 86 3 39 9 Technische Daten...

Page 88: ...enanschluss 2 Anschluss Kanal 1 nur mit CPH6300 S1 3 Anschluss Kanal 2 nur mit CPH6300 S2 4 Anschluss Kanal 1 nur mit CPH6300 S2 Referenz Drucksensor CPT6200 Typ CPH6300 S1 Typ CPH6300 S2 1 3 4 1 2 27...

Page 89: ...ZZZZZ Z Verl ngerungskabel zum Anschuss der Sensoren ca 3 8 m 12 5 ft auf ca 5 m 16 4 ft CPH A 63 ZVZZZZZZZZ Z 2 adriges Anschlusskabel 2 m 6 6 ft mit losen Enden Adern h lsen f r konfigurierbaren Ana...

Page 90: ...90 WIKA Betriebsanleitung Typ CPH6300 DE 14043054 02 08 2017 EN DE...

Page 91: ...91 WIKA Betriebsanleitung Typ CPH6300 DE 14043054 02 08 2017 EN DE...

Page 92: ...Alexander Wiegand SE Co KG Alexander Wiegand Stra e 30 63911 Klingenberg Germany Tel 49 9372 132 0 Fax 49 9372 132 406 info wika de www wika de WIKA subsidiaries worldwide can be found online at www...

Reviews: