19
WIKA operating instructions, model CPH6300
EN
14043054.02 08/2017 EN/DE
Menu
Parameters
Values
Meaning
“MENU”
▶
▲ or ▼
AL
.
Lo
DIF
-19999…
AL.DIF-Hi
Min alarm limit difference (not if
AL
.
diff
=
oFF
)
3)
Sensor-Min ...
AL.Hi
Min alarm limit sensor (not if AL. = oFF) Sensor-
Min is the lower scale range limit of the sensor.
AL
.
Hi
DIF
AL.DIF-Lo…
19999
Max alarm limit difference (not if AL.diff
=
oFF)
3)
AL.Lo ...
Sensor-Max
Max alarm limit sensor (not if
AL.
=
oFF
) Sensor-
Max is
SET
Lo66
Set Logger: Logger function configuration
1)
Selection of the logger function
1)
Fvnc
CYCL
Cyclic: Logger function “cyclic logger”
Stor
Store: Logger function “individual value logger”
oFF
No logger function
CYCL
00:01 ... 60:00
Cycle time for cyclic logger [minutes:seconds]
1)
Lo
.
Po
on/oFF
Low-power logger with very low power supply
(only if cyclic logger with slow measuring rate)
1)
SET
CLOC
Set Clock: Setting of the real-time clock
CLOC
HH:MM
Clock: Set time hours:minutes
YEAr
YYYY
Year: Set year
date
TT.MM
Date: Set time day.month
read
Lo66
rEAd Logg: Reading the individual logger data
See chapter 5.5.1 “Manual records”
1) Menu cannot be selected if there is data in the logger memory. If this is to be changed, the data must first be deleted (see chapter
5.5 “Data logger”).
2) This menu can only be selected if an appropriate sensor is connected to connection 1. When using a second corresponding
sensor on connection 2 then the settings are adopted.
3) Only for 2-channel version
5. Commissioning, operation