Safety instructions and warnings
16/70
Venting
The tank must be vented 3 days after commissioning or after it has been
initially filled – see page 54. After this, venting is unnecessary unless leak-
ages occur and the tank has to be topped up.
The device may no longer be used in the event of water loss (puddles on the
floor). Please inform our customer service department – see page 2.
1.5.6
Operation and maintenance
Error-free operation, maintenance and service are required in order to ensure
that the health and safety of the patients is not endangered during operation
and to ensure that the device is in a fault-free operating condition.
Always abide by the information and recommendations given in this operating
manual. Ensure that the intervals for the inspection, maintenance and service
are complied with. The maintenance work must be carried out by the custom-
er services of the manufacturer or by qualified specialist personnel authorized
by the manufacturer following a check list. Inspection work that has been
carried out correctly can be entered and confirmed on 'Sheet D Maintenance
measures / Service' in the Book of Medical Devices, see page 66.
The current operating hours can be called up in “Default settings”, see
chapter 6.
We would like to point out that the device must be subjected to regular in-
spections by our customer service or an approved specialized company every
12 months (from commissioning) to maintain its proper condition!
Unauthorized modifications and changes to the device and its control unit are
prohibited. Failure to comply with this voids the operating licence!
To avoid hazards, the device must not be used if the mains power cable is
damaged. A damaged mains power cable must be replaced by the manufac-
turer or their customer services, or by trained and qualified specialist person-
nel.
1.5.7
Decommissioning
The device must be disconnected from the power supply in order to tempo-
rarily or permanently decommission it.
You must abide by the legal disposal requirements when permanently de-
commissioning the device.
1.5.8
Storage
Store the devices in a dry, frost-free location with a stable temperature. The
well, pump and the hoses must not contain any water. Once the device has
cooled off, pack it in plastic wrapping to protect it against scratches.
Storage temperature: Between -15 °C and +60 °C